Setup mode allows viewing and modifying the following parameters:
Calibration Alarm Time Out
One-point calibration behaviour
Custom Buffer 1
Custom Buffer 2
Alarm High Limit
Alarm Low Limit
AutoLOG Start Condition
AutoLOG End Condition
AutoLOG Interval
Temperature Unit
Current Date (yyyy.mm.dd)
Current Time (hh:mm)
Printing Language (HI 122 only)
Printer Status (HI 122 only)
Beep Status
Baud Rate (serial communication)
Command prefix (serial communication)
Instrument ID
To enter SETUP mode, press 2nd then SET key, or simply SET from the left
keyboard (HI 120 only), while the instrument is in measurement mode.
Select the desired setup parameter using the ARROW keys.
Press CAL if you want to change the item value. The selected item (e.g.
hour, in setting up the correct time) and “CFM” tag will start blinking.
Press the ARROW keys to change the displayed value or simply use the
numerical keys for all numerical parameters.
Note: To use the numerical keys, press NUM from the left keyboard. The “2nd”
tag will blink. Set the desired number digit by digit. The new introduced
digit is always the last one. All the previous digits will shift to left.
To delete digits press CLR. Last introduced digit will be deleted and
the number will shift to right.
SETUP