All the instruments are factory calibrated for mV.
Hanna’s ORP electrodes are interchangeable and no mV calibration is
needed when they are replaced.
If the mV measurements are inaccurate, mV recalibration should be
performed.
For an accurate recalibration, contact your dealer or the nearest
Hanna Customer Service Center, or follow the instructions below.
A two-point calibration can be performed at 0.0 mV and 1800.0 mV.
Attach to the BNC connector a mV simulator with an accuracy of
±0.1 mV.
With the instrument off, press and hold down the CAL & keys,
then power on the instrument. The CAL tag will appear and the
secondary LCD will show 0.0 mV”.
Set 0.0 mV on the simulator.
When the reading is stable and close to the selected calibration
point, READY tag will appear and CFM” tag will blink.
Press CFM to confirm. The secondary LCD will display “1800 mV”.
Set 1800.0 mV on the simulator.
When the reading is stable and close to the selected calibration
point, READY” tag will appear and CFM” tag will blink.
Press CFM to confirm. The instrument returns to measurement
mode.
Note: If the reading is not close to the selected calibration point,
WRONG” tag will blink. Verify calibration condition or contact
your vendor if you can not calibrate.
mm
mm
m
V CALIBRATIONV CALIBRATION
V CALIBRATIONV CALIBRATION
V CALIBRATION
((
((
(
for technical personnel onlyfor technical personnel only
for technical personnel onlyfor technical personnel only
for technical personnel only
))
))
)