16
RELATIVE RELATIVE
RELATIVE RELATIVE
RELATIVE
mm
mm
m
V CALIBRATIONV CALIBRATION
V CALIBRATIONV CALIBRATION
V CALIBRATION
• Press CAL when the instrument is in RELATIVE mV measurement
mode. The “mV” and “ ” tags will be displayed. Absolute mV is
displayed on the primary LCD and “AbS” message is displayed on the
secondary LCD.
• When the absolute reading is stable and in measurement range, the
instrument asks for confirmation.
• If the reading is out of range, “WRONG” tag will be displayed.
• Press CFM to confirm the absolute value. The instrument will display
“0.0 mV“ on the primary LCD and “rEL” message on the secondary
LCD. In this moment the relative mV offset is equal to absolute mV
reading.
• Use the ARROW keys if you want to change the displayed relative
mV value.
• Press CFM to confirm the relative mV value. The relative mV
offset is displayed on the primary LCD. The instrument returns to
measurement mode.
Note: The relative mV value can be changed only inside the relative mV
offset window (± 2000 mV).