If the meter is in Atc mode and the buffer temperature is out of
0.0  60.0 ºC interval, WRONG ºC tags and the temperature
will be displayed blinking.
For best results choose an EC buffer value close to the sample to be
measured.
In order to minimize any EMC interference, use plastic or glass
beakers.
It is possible to set the cell constant value directly, without following
the calibration procedure. To set the cell constant, enter SETUP
mode and select CEL (see SETUP for details, page 31).
WRONG
CAL