log type (2 chars)
GLDxxx Requests all records for the lot with ID=xxx
The answer string contains:
• Lot header data:
•logging interval (5 char)
•log type (1 char)
•temperature mode (1 char)
•offset (3 char)
•slope (4 char)
•unit (1 char)
Lot record data:
•temperature (3 char)
•value (6 char)
•second value (6 char)
LODxPyyyRequests the yyy
th
pH record logged data when x=1.
LODxMyyy Requests the yyy
th
mV/Rel mV record logged data.
LODxIyyy Requests the yyy
th
ISE record logged data when x=1 (HI 3221)
or when x=2 (HI 3222).
LODxPALL Requests all pH Log on demand when x=1.
LODxMALL Requests all mV/Rel mV Log on demand.
LODxIALLRequests all ISE Log on demand when x=1 (HI 3221)
or when x=2 (HI 3222).
The answer string for each record contains:
The logged mode (2 chars):
xx=10 pH range/0.001 resolution on channel 1
xx=11 pH range/0.01 resolution on channel 1
xx=12 pH range/0.1 resolution on channel 1
xx=13 mV range on channel 1
xx=14 Relative mV range on channel 1
xx=15 ISE range on channel 1 (HI 3221)
xx=20 ISE range on channel 2 (HI 3222)
xx=21 mV range on channel 2 (HI 3222)
xx=22 Relative mV range on channel 2 (HI 3222)
Reading status (1 char): R, O, U
Calculated reading, with sign and decimal point and
exponent (13 chars) - for pH, Rel mV and ISE range
Temperature reading, with sign and two decimal points
(8 chars)
mV reading status (1 char): R, O, U
Vista de pagina 53