mV & Relative mV MEASUREMENTS
mV/ORP MEASUREMENTS
Oxidation-reduction potential (ORP) measurements provide the quantification of the oxidizing or reducing power of the
tested sample.
To correctly perform a redox measurement, the surface of the ORP electrode must be clean and smooth.
DIRECT MEASUREMENT
To measure the mV of a sample using the Direct reading mode:
Press
and then to enter mV Measure mode
(for the selected channel – HI 4212 only).
Select the Direct reading mode (see mV Setup for details, page 32).
Submerge the tip of the ORP electrode (4 cm/1½”) into the
sample to be tested and allow a few seconds for the reading
to stabilize.
The instrument will display the measured mV value on the LCD.
Note: If the reading is out of range, “-----” will be displayed
on the LCD.
DIRECT/AUTOHOLD MEASUREMENT
To measure mV of a sample using the Direct/AutoHold reading mode:
Press
and then to select mV Measure mode
(for the selected channel – HI 4212 only).
Select the Direct/AutoHold reading mode (see mV Setup for
details, page 32).
Submerge the tip of the ORP electrode approximately 4
cm (1½”) into the sample to be tested and allow a few
seconds for the reading to stabilize.
• The measured mV value will be displayed on the LCD. If
pressing
, the “AutoHold” indicator will start blinking
on the LCD until the stability criterion is reached. The mV value
will be frozen on the LCD, along with “AutoHold” indicator.
To return to normal Measure mode press
.
Note: If the reading is out of range, “-----” will be displayed
on the LCD.