To edit the Custom Buffers:
Use
or to select the
Calibration
option.
Press and use or to highlight the
Edit Custom Buffers
option.
For a previous set value, press to set the custom buffer value
to “----” if desired and confirm the setting by pressing ,
otherwise press to edit the selected custom buffer.
While in edit custom buffer menu press to set the
custom buffer value to 7.000 pH and then use or
to set the desired custom buffer value.
Press to exit custom buffer edit menu. If the Saving Confirmation is enabled, press to
accept the modified option, to escape without saving or to return to the editing mode.
Otherwise, the modified option is saved automatically.
Use key to select the next custom buffer to be set or press to return to Calibration options.
Edit Buffer Group
Accessing this option the user can edit the desired group of five pH buffers for automatic buffer recognition
(Automatic Buffer Entry Type). If the Buffer Group already contains five pH buffers, at least one pH buffer has to
be removed in order to add another buffer.
To set the Buffer Group:
Use or to select the
Calibration
option.
Press and use or to highlight the
Edit Buffer Group
option.
Press and use and to choose the
pH buffer to be included in the buffer group.
Press or to add / remove the selected pH
buffer to / from the buffer group.
Press
to return to Calibration options and to save the changes.
90
Make sure the instrument has been calibrated before taking resistivity measurements.
DIRECT MEASUREMENT
To measure the resistivity of a sample using the
Direct
reading
mode:
Press and then to select resistivity mea-
sure mode.
Select the
Direct
reading mode (see
Resistivity
Setup
section).
Proceed as for the conductivity measurement (see
Conductivity
Measurement
section).
DIRECT/AUTOHOLD MEASUREMENT
To measure resistivity of a sample using the
Direct / AutoHold
reading
mode:
Select the
Direct / AutoHold
reading mode (see
Resistivity Setup
section).
Proceed as for the conductivity measurement. (see
Conductivity
Measurement
section)
RESISTIVITY MEASURERESISTIVITY MEASURE
RESISTIVITY MEASURERESISTIVITY MEASURE
RESISTIVITY MEASURE
MENTMENT
MENTMENT
MENT