125124
SPECIFICATIONS
Range 0 to 100 mg/L
Resolution 5 mg/L
Accuracy ±5 mg/L ±3% of reading at 25 °C
Light Source Tungsten lamp with narrow band interference filter @ 466 nm
Method Sulfate is precipitated with barium chloride crystals. Light absorbance of the suspension
is measured.
REQUIRED REAGENTS
Code Description Quantity
HI 93751-0 Indicator reagent 1 packet
REAGENT SETS
HI 93751-01 Reagents for 100 tests
HI 93751-03 Reagents for 300 tests
For other accessories see page 132.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the COLORED OR TURBID SAMPLES procedure on page 17.
• Select the
Sulfate
method using the procedure described
in the
Method Selection
section (see page 12).
• Fill a cuvette with 10 mL of unreacted sample (up to
the mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
Sulfate
SULFATE
10 mL
Silver
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Fill a second cuvette up to the mark with 10 mL of the reacted sample.
• Add 3 drops of Dispersing Agent (HI 93703-51),
replace the cap and invert gently to mix for about
10 seconds.
• Insert the second cuvette into the instrument.
• Press Read to start the reading. The instrument displays the results in mg/L of silver.
INTERFERENCES
Interference may be caused by:
Al
3+
above 30 mg/L Fe
2+
above 1.5 mg/L
Ca
2+
above 1000 mg/L as CaCO
3
Fe
3+
above 10 mg/L
Cd
2+
above 20 mg/L K
+
above 500 mg/L
Cl¯ above 8000 mg/L Mn
2+
above 25 mg/L
Co
2+
above 1.5 mg/L Mg
2+
above 1000 mg/L as CaCO
3
Cr
3+
above 20 mg/L Na
+
above 5000 mg/L
Cr
6+
above 40 mg/L Ni
2+
above 1.5 mg/L
Cu
2+
above 15 mg/L Pb
2+
above 20 mg/L
F¯ above 20 mg/L Zn
2+
above 30 mg/L
10 mL
# 2
# 2