20 21
Free Chlorine
Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.
Remove the cuvette.
Powder reagents procedure
Add the content of one packet of HI 93701-0 DPD reagent.
Replace the cap and shake gently for 20 seconds (or 2 minutes for
of seawater analysis).
Wait for a minute to allow the undissolved reagent to precipitate
and reinsert the cuvette into the instrument.
Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait
for 1 minute and press READ. When the timer ends the meter will perform the reading. The instrument
displays the results in mg/L of free chlorine.
Liquid reagents procedure
To an empty cuvette add 3 drops of HI 93701A-F DPD1
indicator and 3 drops of HI 93701B-F DPD1 buffer. Swirl
gently to mix, and immediately add 10 mL of unreacted
sample. Replace the cap and shake gently again.
x 3x 3
FREE CHLORINEFREE CHLORINE
FREE CHLORINEFREE CHLORINE
FREE CHLORINE
SPECIFICATIONS
Range 0.00 to 2.50 mg/L
Resolution 0.01 mg/L
Accuracy ±0.03 mg/L ±3% of reading
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 525 nm
Method Adaptation of the
EPA DPD method 330.5.
The reaction between free chlorine and the
DPD reagent causes a pink tint in the sample.
REQUIRED REAGENTS
POWDER:
Code Description Quantity
HI 93701-0 DPD 1 packet
LIQUID:
Code Description Quantity
HI 93701A-F DPD1 Indicator 3 drops
HI 93701B-F DPD1 Buffer 3 drops
REAGENT SETS
HI 93701-F Reagents for 300 tests (liquid)
HI 93701-01 Reagents for 100 tests (powder)
HI 93701-03 Reagents for 300 tests (powder)
For other accessories see page 46.
MEASUREMENT PROCEDURE
Select the
Free Chlorine
method using the procedure
described in the
Method Selection
section (see page 11).
Fill the cuvette with 10 mL of unreacted sample (up to
the mark) and replace the cap.
Place the cuvette into the holder and close the lid.
Free Chlorine
FREE CHLORINE
10 mL