Hanna-instruments HI 83205 Manual de usuario Pagina 50

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 76
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 49
50
Fill one graduated mixing cylinder up to the 25 mL
mark with the sample.
Add the content of one packet of HI 93730A-0
molybdenum reagent, close the cylinder and invert it
several times until completely dissolved.
Add the content of one packet of HI 93730B-0
molybdenum reagent to the cylinder, close and invert
it several times until completely dissolved.
Add the content of one packet of HI 93730C
molybdenum reagent to the cylinder, close and
shake it vigorously.
Fill an empty cuvette with 10 mL of sample up to the
mark and replace the cap.
Insert the cuvette into the instrument.
Molybdenum
25 mL
10 mL
Vista de pagina 49
1 2 ... 45 46 47 48 49 50 51 52 53 54 55 ... 75 76

Comentarios a estos manuales

Sin comentarios