1Instruction ManualHI 95734CFree & Total ChlorineHigh Range ISMwww.hannainst.comThis Instrument is inCompliance with the CE Directives24MAN95734R3
10The flashing digits indicate that:1) the month can be set2) the day can be set“Storage”: this prompt appears for 1 second at the end of thecalibrati
11ERROR MESSAGESa) on zero readingThis indicates that the zeroing procedure failed due to a low signal-to-noise ratio. In this case press ZERO again.
12• In order to measure exactly 5 mL of reagent with the 5 mL syringe:(a) push the plunger completely into the syringe and insert the tip into the rea
2All rights are reserved. Reproduction in whole or in part is prohibited without the written consentof the copyright owner, Hanna Instruments Inc., Wo
3Please examine this product carefully. Make sure that the instrument is not damaged. If any damageoccured during shipment, please notify your Dealer.
4The HI 95734 is an auto diagnostic portable microprocessor meter that benefits from Hanna’syears of experience as a manufacturer of analytical instru
5Range Free Cl20.00 to 10.00 mg/LTotal Cl20.00 to 10.00 mg/LResolution 0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 mg/L above 3.50 mg/LPrecision ±0.06 mg/L
6PRINCIPLE OF OPERATIONAbsorption of Light is a typical phenomenon of interaction between electromagnetic radiation andmatter. When a light beam cross
7A microprocessor controlled special tungsten lamp emits radiation which is first optically conditionedand beamed to the sample contained in the cuvet
8KEYPAD DESCRIPTION• ON/OFF: to turn the meter on and off.• ZERO/SET: this is a bi-functional key: press to zero the meter prior to measurement, or ho
9This prompt appears for 1 second each time the instrument isturned on.The parameter code, “734f” for free chlorine or “734t” for totalchlorine, indic