Hanna-instruments HI 96718 Manual de usuario Pagina 15

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 24
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 14
15
VALIDATION PROCEDURE
Alkalinity above 250 mg/L and acidity above 150 mg/L as CaCO
3
require neutralization since the
color could instantly fade. To resolve this, neutralize the sample with diluted HCl or NaOH.
In case of water with hardness greater than 500 mg/L CaCO
3
, shake the sample for approxi-
mately 2 minutes after adding the reagent.
Use the validation procedure to ensure that the instrument is properly calibrated.
Warning: Do not validate the instrument with any standard solutions other than the HANNA CAL
CHECK
TM
Standards, otherwise erroneous results will be obtained.
Turn the meter on by pressing ON/OFF.
• Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
Place the CAL CHECK
TM
Standard
HI 96718-11 Cuvette A into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
Vista de pagina 14
1 2 ... 10 11 12 13 14 15 16 17 18 19 20 ... 23 24

Comentarios a estos manuales

Sin comentarios