5
SPECIFICATIONS
PRECISION AND ACCURACY
Range 0.00 to 2.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.11 mg/L ±5% of reading @ 25°C
Typical EMC Deviation ±0.02 mg/L
Light Source Tungsten lamp
Light Detector Silicon Photocell with narrow band interference filter @ 525 nm
Method Adaptation of the
Standard Methods for the Examination of Water and
Wastwater, 18
th
edition
, EDTA colorimetric method. The reaction
between Mg and reagents causes a violet tint in the sample.
Environment 0 to 50°C (32 to 122°F); max 95% RH non-condensing
Battery Type 1 x 9 volt
Auto-Shut off After 10' of non-use in
measurement mode
;
after 1 hour of non-use in
calibration mode
;
with last reading reminder.
Dimensions 192 x 102 x 67 mm (7.6 x 4 x 2.6")
Weight 290 g (10 oz.).
REQUIRED REAGENTS
Code Description Quantity/test
HI 93719A-0 Mg indicator 0.5 mL
HI 93719B-0 Alkali solution 0.5 mL
HI 93719C-0 EDTA solution 1 drop
HI 96719D-0 EGTA solution 1 drop
PrecisionPrecision
PrecisionPrecision
Precision is how closely repeated measurements
agree with each other. Precision is usually
expressed as standard deviation (SD).
AccuracyAccuracy
AccuracyAccuracy
Accuracy is defined as the nearness of a test
result to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure
explains these definitions.
In a laboratory using a standard solution of 1.00
mg/L magensium hardness and a representative
lot of reagent, an operator obtained with a single
instrument a standard deviation of 0.03 mg/L.