ERRORS AND WARNINGS
The instrument shows clear messages when erroneous condition appears. Messages are also displayed
when the obtained values are outside expected range. The beeper is playing a beep on errors.
a) on zero reading
Light High: There is too much light to perform a measurement.
Please check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement.
Please check the preparation of the zero cuvette.
No Light:The instrument cannot adjust the light level. Please
check that the sample does not contain any debris.
Inverted cuvettes: The sample and the zero cuvette are inverted.
Zero: A zero reading was not taken. Follow the instructions of
the measurement procedure for zeroing the meter.
b) on sample reading