Use the ARROW keys to set the reading on the secondary LCD to that of
the hot water.
When the reading is stable and close to the selected
calibration point, READY tag will appear and CFM
tag will blink.
Press CFM to confirm. The instrument returns to
measurement mode.
Note: If the reading is not close to the selected calibration point, WRONG
tag will blink. Change the temperature probe and restart calibration.
C C