25
Data transmission from the instrument to the PC can be done with the HI
92000 Windows
®
compatible software (optional). HI 92000 also offers
graphing and on-line help feature.
Data can be exported to the most popular spreadsheet programs for further
analysis.
To connect your instrument to a PC, use the optional Hanna HI 920010 cable
connector. Make sure that your instrument is switched off and plug one
connector to the instrument RS232C socket and the other into the serial port
of your PC.
Note: Other cables than HI 920010 may use a different configuration.
In this case communication between instrument and PC may not
be possible.
If you are not using Hanna Instruments HI 92000 software,
please see the following instructions.
SENDING COMMANDS FROM PC
It is also possible to remotely control the instrument with any terminal
program. Use HI 920010 cable to connect the instrument to a PC, start the
terminal program and set the communication options as follows: 8, N, 1, no
flow control.
COMMAND TYPES
To send a command to the instrument the scheme is:
<command prefix> <command> <CR>
where: <command prefix> is a selectable ASCII character
between 0 and 47.
<command> is the command code (3 characters).
Note: Either small or capital letters can be used.
SIMPLE COMMANDS
PHR Sets the range to pH 0.001 resolution
MVR Sets the range to mV (HI 113 only)
MRS Sets the range to relative mV (HI 113 only)
RNG Is equivalent to pressing RANGE
CAL Is equivalent to pressing CAL
CFM Is equivalent to pressing CFM/GLP
UPC Is equivalent to pressing the UP arrow key
DWC Is equivalent to pressing the DOWN arrow key
MEM Is equivalent to pressing MEM
MRR Is equivalent to pressing MR
SET Is equivalent to pressing SET
PC INTERFACE (HI 112 AND HI 113)