- or the number of days until the calibration alarm will be displayed.
- or if expired (7 days ago).
The instrument ID.
Notes: If a one-point calibration is performed after a two-point
calibration, the instrument will keep the old slope.
If “no bUF” message appears on the LCD, the instrument
informs you that calibration was performed at only one
point.
Press GLP at any moment and the instrument will return to
measurement mode.
If calibration has not been performed, the instrument
displays “no CAL” message blinking.
LAST RELATIVE mV CALIBRATION DATA
Last Relative mV calibration data is stored automatically after a successful
calibration.
To view the Relative mV calibration data, press GLP while in Relative mV
measurement mode.
The instrument will display the Relative mV GLP information and the
time (hh:mm:ss) of the last Rel mV calibration.
Use the ARROW keys to scroll through the calibration data:
The Relative mV calibration date.
The Relative mV calibration offset.
The instrument ID.
Notes: Press GLP to return to measurement mode.
If calibration has not been performed, the instrument
displays “no CAL” message blinking.