
30
ONE-POINT CALIBRATION BEHAVIOR
Press CAL when “1 Pnt” message is displayed on the secondary LCD.
One of the two options (“Pnt” or “OFFS”) and “CFM” will start
blinking (see pH CALIBRATION PROCEDURE for details, page 9).
Press the ARROW keys to toggle between “Pnt” and “OFFS” options.
Press CFM to save the behavior for one-point calibration.
Press CAL to escape without saving.
pH RESOLUTION
Press CAL when “rES” message is displayed on the secondary LCD. The
set resolution (“0.1”, “0.01” or “0.001”) and “CFM” will start blinking.
Press the ARROW keys to toggle between 0.1, 0.01 and 0.001 options.
Press CFM to save the modified value.
Press CAL to escape without saving.
EGRAHCNOI sepytNOI
2– )snoinatnelavid( S
1– )snoinatnelavonom( OlC,NCS,NC,I,rB,lC,F
4
ON,
3
,
HN
3
1 )snoitactnelavonom( HN,gA,K,aN
4
OC,
2
2 )snoitactnelavid( bP,uC,dC,aB,aC,gM
Fdnu noidenifednu
ION CHARGE
Press CAL when “IonC” is displayed. The ion charge (“UndF”, “–2”,
“–1”, “1” or “2”) and “CFM” will start blinking.
Press the ARROW
keys to change the ion charge.
Press CFM to save the modified ion charge.
Press CAL to escape without saving.
Note: To select the right ion charge, different ion types and their charge
are presented in the table below.
Comentarios a estos manuales