Code Valid Values Default PW
14 Relay 1 deviation (D1) 0.5 to 10% f.s. 1% f.s. no
15 Relay 1 reset time 0.1 to 999.9 minutes 999.9 no
16 Relay 1 rate time 0.0 to 999.9 minutes 0.0 no
21 Relay 2 mode (M2) same as relay 1 0 no
22 Relay 2 setpoint (S2) 0.5 to 99.5% full scale 75% f.s. no
23 Relay 2 hysteresis (H2) 0 to 5% f.s. 1% f.s. no
24 Relay 2 deviation (D2) 0.5 to 10% f.s. 1% f.s. no
25 Relay 2 reset time 0.1 to 999.9 minutes 999.9 no
26 Relay 2 rate time 0.0 to 999.9 minutes 0.0 no
30 Relay 3 high alarm (HA) 0.5 to 99.5% full scale 95% f.s. no
HA-HysLA+Hys,Hys=1.5%f.s.,HAS1 or HAS2
31 Relay 3 low alarm (LA) 0.5 to 99.5% full scale 5% f.s. no
LA+HysHA-Hys,Hys=1.5%f.s.,LAS1 or LAS2
32 Proportional control 1 to 30 min 5 no
mode period
33 Maximum relay ON time 10 to 9999 min 60 no
(after which an alarm mode is entered)
34 Alarm mask time 00:00 to 30:00 00:00 no
40 Analog output selection 0: 0-1mA 2 no
1: 0-20 mA
2: 4-20 mA
3: 0-5 VDC
4: 1-5 VDC
5: 0-10 VDC
41 Analog output 0 to 100% full scale 0 no
lower limit (O_VARMIN) (O_VARMIN O_VARMAX - 5% f.s.)
42 Analog output 0 to 100% full scale 100% f.s. no
upper limit (O_VARMAX) (O_VARMIN O_VARMAX - 5% f.s.)
After confirmation, the selected parameter
is displayed. The user can scroll through
the parameters by pressing CFM.
In order to directly set another param-
eter, press SETUP again and enter the
code or scroll to it by pressing CFM.
The following table lists the setup codes along with the descrip-
tion of the specific setup items, their valid values and whether
password is required to view that item (“PW” column):
Code Valid Values Default PW
00 Factory ID 0 to 9999 0000 no
01 Process ID 0 to 99 00 no
02 Control enable/disable 0: C.M. disabled 0 no
1: C.M. enabled
03 Range 1: 0.0-199.9 μS (or 100.0 ppm) 4 no
(depends on model) 2: 0-1999 μS (or 1000 ppm)
3: 0.00-19.99 mS (or 10.00 ppt)
4: 0.0-199.9 mS (or 100.0 ppt)
04 Reference Temperature 20°C or 25°C 25°C no
05 Temperature Coefficient 0.00 to 10.00 %/°C 2.00 no
06 Input Selection 0: conductivity probe 0 no
1: 4-20 mA input signal
07 Temperature ATC: Automatic ATC no
compensation User: Manual
08 TDS Factor (HI710 only) 0.00 to 1.00 0.50 no
11 Relay 1 mode 0: disabled 0 no
(M1) 1: ON-OFF high setpoint
2: ON-OFF low setpoint
3: PID, high setpoint
4: PID, low setpoint
12 Relay 1 setpoint (S1) 0.5 to 99.5% full scale 25% f.s. no
13 Relay 1 hysteresis (H1) 0 to 5% f.s. 1% f.s. no