Setup Item Description Valid Values Default
ERROR CONFIGURATION
E.21 Temperature level Alarm relay -------------> On
error (21) 22 mA fault current----> On
3.6 mA fault current---> Off
SMS sending -----------> Off
E.40 Digital transmitter Alarm relay -------------> On
error (40) 22 mA fault current----> On
3.6 mA fault current---> Off
Hold mode --------------> On
SMS sending -----------> Off
E.50 GSM/Modem Alarm relay -------------> On
module error (50) 22 mA fault current----> On
3.6 mA fault current---> Off
E.60 Temp. compensation Alarm relay -------------> On
error (60) 22 mA fault current----> On
3.6 mA fault current---> Off
SMS sending -----------> Off
E.61 Temperature out of Alarm relay -------------> On
concentration table 22 mA fault current----> On
(61) 3.6 mA fault current---> Off
SMS sending -----------> Off
E.62 Conductivity out of Alarm relay -------------> On
concentration table 22 mA fault current----> On
(62) 3.6 mA fault current---> Off
Auto-cleaning ----------> Off
SMS sending -----------> Off
E.63 Concentration out Alarm relay -------------> On
of user concentration 22 mA fault current----> On
table (63) 3.6 mA fault current---> Off
SMS sending -----------> Off
E.90 Power reset (90) Alarm relay -------------> Off
22 mA fault current----> On
3.6 mA fault current---> Off
SMS sending -----------> Off
Setup Item Description Valid Values Default
ERROR CONFIGURATION
E.01 Low alarm (01) Alarm relay -------------> On
22 mA fault current----> Off
3.6 mA fault current---> On
Auto cleaning ----------> Off
SMS sending -----------> Off
E.02 Max relay ON time Alarm relay -------------> On
error (02) 22 mA fault current----> On
3.6 mA fault current---> Off
Auto cleaning ----------> Off
SMS sending -----------> Off
E.03 Life check error (03) Alarm relay -------------> On
(see note 9) 22 mA fault current----> On
3.6 mA fault current---> Off
Hold mode --------------> On
Auto cleaning ----------> Off
SMS sending -----------> Off
E.10 Conductivity input Alarm relay -------------> On
overflow (10) 22 mA fault current----> On
3.6 mA fault current---> Off
Hold mode --------------> Off
Auto cleaning ----------> Off
SMS sending -----------> Off
E.12 Calibration timeout Alarm relay -------------> Off
(12) 22 mA fault current----> Off
3.6 mA fault current---> Off
SMS sending -----------> Off
E.20 Broken temperature Alarm relay -------------> On
probe (20) 22 mA fault current----> On
3.6 mA fault current---> Off
Hold mode --------------> Off
SMS sending -----------> Off