Instruction Manual
HI 83200
Multiparameter Bench
Photometer
for Laboratories
www.hannainst.com
This Instrument is in
Compliance with the CE Directives
Vista de pagina 0
1 2 3 4 5 6 ... 66 67

Indice de contenidos

Pagina 1 - HI 83200

1Instruction ManualHI 83200Multiparameter BenchPhotometerfor Laboratorieswww.hannainst.comThis Instrument is inCompliance with the CE Directives

Pagina 2 - TABLE OF CONTENTS

18 19• Remove the cap and fill the Demineralizer Bottle with tapwater.• Replace the cap and shake gently for at least 2 minutes.• Open the upper part

Pagina 3 - PRECISION AND ACCURACY

20 21AluminumALUMINUMSPECIFICATIONSRange 0.00 to 1.00 mg/LResolution 0.01 mg/LAccuracy ±0.02 mg/L ±4% of readingTypical EMC ±0.01 mg/LDeviationLight S

Pagina 4 - FUNCTIONAL DESCRIPTION

22 23• Add the content of one packet of HI 93712C-0 Bleachingpowder to one of the two cuvettes. Replace the cap andshake vigorously until completely d

Pagina 5 - USING CUVETTES

24 25Ammonia MRSPECIFICATIONSRange 0.00 to 10.00 mg/LResolution 0.01 mg/LAccuracy ±0.05 mg/L ±5% of readingTypical EMC ±0.01 mg/LDeviationLight Source

Pagina 6 - HEALTH & SAFETY

26 27• Remove the cuvette.• Add 4 drops of HI 93715A-0 First Reagent (6 drops for seawateranalysis). Replace the cap and mix the solution.• Add 4 drop

Pagina 7 - OPERATIONAL GUIDE

28 29Ammonia LRBromineBROMINESPECIFICATIONSRange 0.00 to 8.00 mg/LResolution 0.01 mg/LAccuracy ±0.08 mg/L ±3% of readingTypical EMC ±0.01 mg/LDeviatio

Pagina 8 - SETUPSETUP

30 31• Remove the cuvette and add the content of one packet of HI 93716-0DPD reagent. Replace the cap and shake gently for about 20 seconds todissolve

Pagina 9 - HELP MODE

32 33FREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINESPECIFICATIONSRange 0.00 to 2.50 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3%

Pagina 10

34 35• Reinsert the cuvette into the instrument.• Press READ to start the reading. The instrument displays the resultsin mg/L of free chlorine.INTERFE

Pagina 11 - ALUMINUM

36 37Total Chlorine• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.Powder reagen

Pagina 12 - ALKALINITY

2 3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If any damageoccurred during shipment,

Pagina 13 - AMMONIA MEDIUM RANGE

38 39Chlorine DioxideCHLORINE DIOXIDESPECIFICATIONSRange 0.00 to 2.00 mg/LResolution 0.01 mg/LAccuracy ±0.10 mg/L ±5% of readingTypical EMC ±0.01 mg/L

Pagina 14 - AMMONIA LOW RANGE

40 41Chlorine Dioxide• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Fill another cuvette with 10 m

Pagina 15 - MEASUREMENT PROCEDURE

42 43• Reinsert the cuvette into the instrument.• Press TIMER and the display will show the countdown prior to themeasurement or, alternatively, wait

Pagina 16

44 45• Reinsert the cuvette into the instrument.• Press TIMER and the display will show the countdown prior to themeasurement or, alternatively, wait

Pagina 17 - FREE CHLORINE

46 47Color of Water# 3• Fill the second cuvette up to the mark with unfiltered sample andreplace the cap. This is the apparent color.• Filter 10 mL of

Pagina 18

48 49Copper HR• Remove the cuvette.• Add the content of one packet of HI 93702-0 Bicinchoninate. Replacethe cap and shake gently for about 15 seconds.

Pagina 19 - TOTAL CHLORINE

50 51Copper LR• Remove the cuvette.• Add the content of one packet of HI 93747-0 Bicinchoninate.Replace the cap and shake gently for about 15 seconds.

Pagina 20 - CHLORINE DIOXIDE

52 53CyanideNote: Pay attention to the way the spoon is filled:- do not press the powder;- do not overfill it.• Place the HDPE plastic stopper and cap

Pagina 21 - SAMPLING PROCEDURE

54 55Cyanuric AcidCYANURIC ACIDSPECIFICATIONSRange 0 to 80 mg/LResolution 1 mg/LAccuracy ±1 mg/L ±15% of readingTypical EMC ±1 mg/LDeviationLight Sour

Pagina 22 - CHROMIUM VI HIGH RANGE

56 57FluorideFLUORIDESPECIFICATIONSRange 0.00 to 2.00 mg/LResolution 0.01 mg/LAccuracy ±5% of readingTypical EMC ±0.01 mg/LDeviationLight Source Tungs

Pagina 23 - CHROMIUM VI LOW RANGE

4 5Where:-log I/Io=Absorbance (A)Io=intensity of incident light beamI =intensity of light beam after absorptionελ=molar extinction coefficient at wave

Pagina 24 - COLOR OF WATER

58 59Hardness CaCALCIUM HARDNESSSPECIFICATIONSRange 0.00 to 2.70 mg/LResolution 0.01 mg/LAccuracy ±0.11 mg/L ±5% of readingTypical EMC ±0.01 mg/LDevia

Pagina 25 - COPPER HIGH RANGE

60 61Hardness CaNote: This test will detect any calcium contamination in the beaker, measuring syringes or sample cells. Totest cleanliness, repeat th

Pagina 26 - COPPER LOW RANGE

62 63Hardness Mg• Fill both cuvettes up to the 10 mL mark.• Add 1 drop of HI 93719C-0 EDTA solution to one cuvette (# 1),replace the cap and invert th

Pagina 27

64 65HydrazineHYDRAZINESPECIFICATIONSRange 0 to 400 µg/LResolution 1 µg/LAccuracy ±4% of full scaleTypical EMC ±2 µg/LDeviationLight Source Tungsten l

Pagina 28 - CYANURIC ACID

66 67IODINESPECIFICATIONSRange 0.0 to 12.5 mg/LResolution 0.1 mg/LAccuracy ±0.1 mg/L ±5% of readingTypical EMC ±0.1 mg/LDeviationLight Source Tungsten

Pagina 29 - FLUORIDE

68 69Iron HRIRON HIGH RANGESPECIFICATIONSRange 0.00 to 5.00 mg/LResolution 0.01 mg/LAccuracy ±0.04 mg/L ±2% of readingTypical EMC ±0.01 mg/LDeviationL

Pagina 30 - CALCIUM HARDNESS

70 71Iron LRIRON LOW RANGESPECIFICATIONSRange 0 to 400 µg/LResolution 1 µg/LAccuracy ±10 µg/L ±8% of readingTypical EMC ±1 µg/LDeviationLight Source T

Pagina 31 - SAMPLE DILUTION

72 73Iron LRINTERFERENCESInterference may be caused by:Cadmium above 4.0 mg/LChromium3+ above 0.25 mg/LChromium6+ above 1.2 mg/LCobalt above 0.05 mg/L

Pagina 32 - MAGNESIUM HARDNESS

74 75Magnesium• Remove the cuvette.• Using the other 1 mL syringe, add to the cuvette exactly 0.5 mL of sample.Note: Do not mix up the two syringes!•

Pagina 33 - HYDRAZINE

76 77Manganese HR• Add the content of one packet of HI 93709B-0 Sodium Periodatereagent. Replace the cap and shake gently until completely dissolved.•

Pagina 34

6 7INSTRUMENT DESCRIPTION1) Cuvette Lid2) Liquid Crystal Display (LCD).3) Splash proof keypad.4) ON/OFF power switch5) Power input connector6) USB con

Pagina 35 - IRON HIGH RANGE

78 79Manganese LR• Add 0.2 mL of the HI 93748B-0 Alkaline-cyanide reagentsolution to each cuvette, replace the caps and shake gently.• Add 1 mL of the

Pagina 36 - 30"

80 81 MolybdenumMOLYBDENUMSPECIFICATIONSRange 0.0 to 40.0 mg/LResolution 0.1 mg/LAccuracy ±0.3 mg/L ±5% of readingTypical EMC ±0.1 mg/LDeviationLight

Pagina 37

82 83• Fill one graduated mixing cylinder up to the 25 mLmark with the sample.• Add the content of one packet of HI 93730A-0molybdenum reagent, close

Pagina 38 - MAGNESIUM

84 85Nickel HR• Reinsert the cuvette into the instrument.• Press TIMER and the display will show the countdown priorto the measurement or, alternative

Pagina 39 - MANGANESE HIGH RANGE

86 87Nickel LR• Press TIMER and the display will show a countdown prior to adding reagent C or, alternatively, waitfor 15 minutes. Add one packet of H

Pagina 40 - MANGANESE LOW RANGE

88 89NitrateSPECIFICATIONSRange 0.0 to 30.0 mg/LResolution 0.1 mg/LAccuracy ±0.5 mg/L ±10% of readingTypical EMC ±0.1 mg/LDeviationLight Source Tungst

Pagina 41 - MOLYBDENUM

90 91Nitrite HRNITRITE HIGH RANGESPECIFICATIONSRange 0 to 150 mg/LResolution 1 mg/LAccuracy ±4 mg/L ±4% of readingTypical EMC ±1 mg/LDeviationLight So

Pagina 42

92 93SPECIFICATIONSRange 0.00 to 0.35 mg/LResolution 0.01 mg/LAccuracy ±0.02 mg/L ±4% of readingTypical EMC ±0.01 mg/LDeviationLight Source Tungsten l

Pagina 43 - NICKEL HIGH RANGE

94 95Dissolved OxygenSPECIFICATIONSRange 0.0 to 10.0 mg/LResolution 0.1 mg/LAccuracy ±0.4 mg/L ±3% of readingTypical EMC ± 0.1 mg/LDeviationLight Sour

Pagina 44 - NICKEL LOW RANGE

96 97OZONESPECIFICATIONSRange 0.00 to 2.00 mg/LResolution 0.01 mg/LPrecision ±0.03 mg/L @ 1.00 mg/LTypical EMC ±0.01 mg/LDeviationLight Source Tungste

Pagina 45

8 9USING LIQUID AND POWDER REAGENTS• Proper use of the dropper:(a) for reproducible results, tap the dropper on the table for several times and wipe

Pagina 46 - NITRITE HIGH RANGE

98 99• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.• Add the content of one

Pagina 47 - NITRITE LOW RANGE

100 101pH• Reinsert the cuvette into the instrument.• Press the READ key to start the reading. The instrument displaysthe pH value.pHpHSPECIFICATIONSR

Pagina 48 - DISSOLVED OXYGEN

102 103Phosphate HR• Add the content of one packet of HI 93717B-0 PhosphateHR Reagent B to the cuvette. Replace the cap and shakegently until complete

Pagina 49

104 105Phosphate LR• Reinsert the cuvette into the instrument.• Press TIMER and the display will show the countdown prior to themeasurement or, altern

Pagina 50

106 107Phosphorus• Add the content of one packet of HI 93706B-0 PhosphorusReagent B (Amino Acid) to the cuvette. Replace the capand shake gently until

Pagina 51

108 109• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette and add the content of

Pagina 52 - PHOSPHATE HIGH RANGE

110 111Potassium MRPOTASSIUM MEDIUM RANGESPECIFICATIONSRange 10 to 100 mg/LResolution 2.5 mg/LAccuracy ±15 mg/L ±7% of readingTypical EMC ±2.5 mg/LDev

Pagina 53 - PHOSPHATE LOW RANGE

112 113Potassium LRPOTASSIUM LOW RANGESPECIFICATIONSRange 0.0 to 20.0 mg/LResolution 0.5 mg/LAccuracy ±3.0 mg/L ±7% of readingTypical EMC ±0.5 mg/LDev

Pagina 54 - PHOSPHORUS

114 115• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette and add the content of

Pagina 55 - POTASSIUM HIGH RANGE

116 117Silica• Place the cuvette into the holder and close the lid.• Press ZERO key. The display will show “-0.0-” when themeter is zeroed and ready f

Pagina 56 - POTASSIUM MEDIUM RANGE

10 11Pay attention to push the cuvette completely down in the holder.• In order to avoid reagent leaking and to obtain more accurate measure-ments, cl

Pagina 57 - POTASSIUM LOW RANGE

118 119Silver• Add exactly 1.0 mL of HI 93737B-0 Buffer reagent to the second beaker(the sample) and swirl gently to mix.• Press TIMER and the display

Pagina 58

120 121SPECIFICATIONSRange 0 to 150 mg/LResolution 5 mg/LAccuracy ±5 mg/L ±3% of readingLight Source Tungsten lamp with narrow band interference filte

Pagina 59

122 123ZincZINCSPECIFICATIONSRange 0.00 to 3.00 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of readingTypical EMC ±0.01 mg/LDeviationLight Source

Pagina 60

124 125The instrument shows clear warning messages when erroneous conditions appear and when measured values areoutside the expected range. These mess

Pagina 61

126 127Description Range MethodAluminum 0.00 to 1.00 mg/L AluminonAlkalinity 0 to 500 mg/L ColorimetricAmmonia MR 0.00 to 10.00 mg /L NesslerAmmonia L

Pagina 62 - REQUIRED REAGENT

128 129HI 93721-01 100 iron HR testsHI 93721-03 300 iron HR testsHI 93722-01 100 cyanuric acid testsHI 93722-03 300 cyanuric acid testsHI 93723-01 100

Pagina 63 - ERRORS AND WARNINGS

130 131WARRANTYAll Hanna Instruments meters are warranted for two years against defects in workmanship and materialswhen used for its intended purpos

Pagina 64 - DATA MANAGEMENT

132Australia:Tel. (03) 9769.0666 • Fax (03) 9769.0699e-mail: [email protected]:Tel. (10) 88570068 • Fax (10) 88570060e-mail: hannachina@v

Pagina 65 - ACCESSORIES

12 13POWER CONNECTION AND BATTERY MANAGEMENTThe meter can be powered from an AC/DC adapter (included) or from the built-in rechargeable battery.Note:

Pagina 66 - HANNA LITERATURE

14 15Date / TimeThis option is used to set the instrument’s date and time.Press Modify functional key to change the date/time.Press the WX functional

Pagina 67 - MAN83200 05/08

16 17HI 83200 offers an interactive contextual help mode that assists the user at any time.To access help screens press HELP.The instrument will displ