Ammonia MR
SPECIFICATIONS
Range 0.00 to 10.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.05 mg/L ±5% of reading
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 420 nm
Method Adaptation of the
ASTM Manual of Water and Environmental Technology, D1426-92
,
Nessler method. The reaction between ammonia and reagents causes a yellow tint in
the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93715A-0 First Reagent 4 drops (6 drops for seawater)
HI 93715B-0 Second Reagent 4 drops (10 drops for seawater)
REAGENT SETS
HI 93715-01 Reagents for 100 tests
HI 93715-03 Reagents for 300 tests
For other accessories see page 73.
MEASUREMENT PROCEDURE
Select the
Ammonia MR
method using the procedure described in
the
Method Selection
section (see page 12).
Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
Place the cuvette into the holder and close the lid.
Press ZERO key. The display will show “-0.0-” when the meter is
zeroed and ready for measurement.
AMMONIA MEDIUM RANGE
Aluminum
Add the content of one packet of HI 93712C-0 Bleaching
Powder to one of the two cuvettes. Replace the cap and
shake vigorously until completely dissolved. This is the blank.
Place the blank into the holder and close the lid.
Press TIMER and the display will show the countdown prior to
zeroing the blank. Alternatively wait for 15 minutes and then press
ZERO. The display will show "-0.0-" when the meter is zeroed and
ready for measurement.
Remove the blank and insert the other cuvette into the instrument.
Press the READ key and the meter will perform the reading. The
instrument displays the results in mg/L of aluminum.
Press the or to access the second level of functions.
Press the Chem Frm functional key to convert the result in mg/L of Al
2
O
3
.
Press the or to go back to the measurement screen.
INTERFERENCES
Interference may be caused by:
Iron above 20 mg/L Alkalinity above 1000 mg/L
Phosphate above 50 mg/L Fluoride must be absent
# 1