Reinsert the cuvette into the instrument.
Press TIMER and the display will show the countdown prior
to the measurement or, alternatively, wait for 6 minutes
and press READ. When the timer ends the meter will perform
the reading. The instrument displays concentration in mg/L
of nitrite.
Press the or to access the second level of functions.
Press the Chem Frm functional key to convert the result in mg/L of nitrogen-nitrite (NO
2
-N) and sodium
nitrite (NaNO
2
).
Press the or to go back to the measurement screen.
INTERFERENCES
Interference may be caused by the following ions:
ferrous, ferric, cupric, mercurous, silver, antimonious, bismuth, auric, lead, metavanadate and chloroplatinate.
Strongly reducing and oxidizing reagents.
High levels of nitrate (above 100 mg/L) could yield falsely high readings due to a minute amount of
reduction to nitrite that could occur at these levels.
Nitrite LR
SPECIFICATIONS
Range 0.00 to 0.35 mg/L
Resolution 0.01 mg/L
Accuracy ±0.02 mg/L ±4% of reading
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 525 nm
Method Adaptation of the
EPA Diazotization method 354.1
. The reaction between nitrite and
the reagent causes a pink tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93707-0 Powder reagent 1 packet
REAGENT SETS
HI 93707-01 Reagents for 100 tests
HI 93707-03 Reagents for 300 tests
For other accessories see page 73.
MEASUREMENT PROCEDURE
Select the
Nitrite LR
method using the procedure described
in the
Method Selection
section (see page 12).
Fill the cuvette up to the mark with 10 mL of unreacted
sample (up to the mark) and replace the cap.
Place the cuvette into the holder and close the lid.
Press ZERO key. The display will show “-0.0-” when the
meter is zeroed and ready for measurement.
Remove the cuvette.
Add the content of one packet of HI 93707-0 reagent.
Replace the cap and shake gently for about 15 seconds.
Nitrite LR
NITRITE LOW RANGE
10 mL