Hanna-instruments HI 83206 Manual de usuario Pagina 41

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 76
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 40
41
Molybdenum
Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 5 minutes and press Read. When the timer ends the meter will perform the reading. The instrument
displays concentration in mg/L of molybdenum.
Press or to access the second level functions.
Press the Chem Frm key to convert the result in mg/L of molybdate (MoO
4
) and sodium molybdate
(Na
2
MoO
4
).
Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Aluminum above 50 mg/L
Chromium above 1000 mg/L
Copper above 10 mg/L
Iron above 50 mg/L
Nickel above 50 mg/L
Nitrite, as NO
2
¯
Sulfate above 200 mg/L
Highly buffered samples or with extreme pH may exceed the buffering capacity of the reagents.
Vista de pagina 40
1 2 ... 36 37 38 39 40 41 42 43 44 45 46 ... 75 76

Comentarios a estos manuales

Sin comentarios