1Instruction ManualHI 83209Multiparameter BenchPhotometerfor Educationwww.hannainst.com
10• Pay attention to push the cuvette completely down in the holder and to align the white point on thecap to the indexing mark on the meter.• In orde
11METHOD REFERENCE TABLEMethod Method Pagedescription1 Ammonia MR 172 Ammonia LR 193 Free Chlorine 214 Total Chlorine 245 Chromium VI HR 276 Chromium
12POWER CONNECTION AND BATTERY MANAGEMENTThe meter can be powered from an AC/DC adapter (included) or from the built-in rechargeable battery.Note: Alw
13DATA MANAGEMENTThe instrument features a data log function to help you keep track of all your analysis. The data log canhold 200 individual measurem
14In the Setup mode the instrument’s parameters can be changed. Some parameters affect the measuringsequence and others are general parameters that ch
15Date / TimeThis option is used to set the instrument’s date and time.Press the Modify key to change the date/time.Press the W X functional keys to h
16HI 83209 offers an interactive contextual help mode that assists the user at any time.To access the help screens press HELP.The instrument will disp
17Ammonia MRSPECIFICATIONSRange 0.00 to 10.00 mg/LResolution 0.01 mg/LAccuracy ±0.05 mg/L ±5% of reading at 25 °CTypical EMC ±0.01 mg/LDeviationLight
18• Remove the cuvette.• Add 4 drops of HI 93715A-0 First Reagent (6 drops forseawater analysis). Replace the cap and mix the solution.• Add 4 drops o
19Ammonia LRSPECIFICATIONSRange 0.00 to 3.00 mg/LResolution 0.01 mg/LAccuracy ±0.04 mg/L ±4% of reading at 25 °CTypical EMC ±0.01 mg/LDeviationLight S
2Dear Customer,Thank you for choosing a Hanna product. Please read this instruction manual carefully before using theinstrument. This manual will prov
20Ammonia LR• Remove the cuvette.• Add 4 drops of HI 93700A-0 First Reagent (6 drops forseawater analysis). Replace the cap and mix the solution.• Add
21FREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINEFREE CHLORINESPECIFICATIONSRange 0.00 to 2.50 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of
22Free Chlorine• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.Powder reagen
23• Reinsert the cuvette into the instrument.• Press Read to start the reading. The instrument displays the results in mg/L of free chlorine.INTERFER
24Total ChlorineTOTAL CHLORINESPECIFICATIONSRange 0.00 to 3.50 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of reading at 25 °CTypical EMC ±0.01 mg
25Total Chlorine• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.• Remove the cuvette.Powder reage
26• Press Timer and the display will show the countdown prior to the measurement or, alternatively, waitfor 2 minutes and 30 seconds and press Read. W
27Chromium VI HRCHROMIUM VI HIGH RANGESPECIFICATIONSRange 0 to 1000 μg/LResolution 1 μg/LAccuracy ±5 μg/L ±4% of reading at 25 °CTypical EMC ±1 μg/LDe
28• Reinsert the cuvette into the instrument.• Press Timer and the display will show the countdown prior to the measurement or, alternatively, waitfor
29Chromium VI LRCHROMIUM VI LOW RANGESPECIFICATIONSRange 0 to 300 μg/LResolution 1 μg/LAccuracy ±1 μg/L ±4% of reading at 25 °CTypical EMC ±1 μg/LDevi
3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If any damageoccurred during shipment, pl
30• Reinsert the cuvette into the instrument.• Press Timer and the display will show the countdown prior to the measurement or, alternatively, waitfor
31Color of WaterCOLOR OF WATERSPECIFICATIONSRange 0 to 500 PCU (Platinum Cobalt Units)Resolution 1 PCUAccuracy ±10 PCU ±5% of reading at 25 °CTypical
32Color of Water• Fill the second cuvette up to the mark with unfiltered sample andreplace the cap. This is the apparent color.• Filter 10 mL of sampl
33Copper HRCOPPER HIGH RANGESPECIFICATIONSRange 0.00 to 5.00 mg/LResolution 0.01 mg/LAccuracy ±0.02 mg/L ±4% of reading at 25 °CTypical EMC ±0.01 mg/L
34Copper HR• Remove the cuvette.• Add the content of one packet of HI 93702-0 Bicinchoninate.Replace the cap and shake gently for about 15 seconds.• R
35Copper LRCOPPER LOW RANGESPECIFICATIONSRange 0 to 1000 μg/LResolution 1 μg/LAccuracy ±10 μg/L ±5% of reading at 25 °CTypical EMC ±1 μg/LDeviationLig
36Copper LR• Remove the cuvette.• Add the content of one packet of HI 95747-0Bicinchoninate. Replace the cap and shake gently forabout 15 seconds.• Re
37NitrateSPECIFICATIONSRange 0.0 to 30.0 mg/LResolution 0.1 mg/LAccuracy ±0.5 mg/L ±10% of reading at 25 °CTypical EMC ±0.1 mg/LDeviationLight Source
38Nitrate• Replace the cap and immediately shake vigorously upand down for exactly 10 seconds. Continue to mix byinverting the cuvette gently for 50 s
39Nitrite HRNITRITE HIGH RANGESPECIFICATIONSRange 0 to 150 mg/LResolution 1 mg/LAccuracy ±4 mg/L ±4% of reading at 25 °CTypical EMC ±1 mg/LDeviationLi
4PRINCIPLE OF OPERATIONAbsorption of light is a typical phenomenon of interaction between electromagnetic radiation and matter.When a light beam cross
40Nitrite HR• Add the content of one packet of HI 93708-0 reagent.Replace the cap and shake gently until completely dissolved.• Reinsert the cuvette i
41SPECIFICATIONSRange 0.00 to 1.15 mg/LResolution 0.01 mg/LAccuracy ±0.06 mg/L ±4% of reading at 25 °CTypical EMC ±0.01 mg/LDeviationLight Source Tung
42• Reinsert the cuvette into the instrument.• Press Timer and the display will show the countdown prior to the measurement or, alternatively, waitfor
43Dissolved OxygenSPECIFICATIONSRange 0.0 to 10.0 mg/LResolution 0.1 mg/LAccuracy ±0.4 mg/L ±3% of reading at 25 °CTypical EMC ± 0.1 mg/LDeviationLigh
44Dissolved Oxygen• Let the sample stand and the flocculent agent will start to settle.• After approximately 2 minutes, when the upper half of the bot
45pHpHSPECIFICATIONSRange 6.5 to 8.5 pHResolution 0.1 pHAccuracy ±0.1 pH at 25 °CTypical EMC ±0.1 pHDeviationLight Source Tungsten lamp with narrow ba
46pH• Remove the cuvette and add 5 drops of HI 93710-0Phenol Red Indicator. Replace the cap and mix thesolution.• Reinsert the cuvette into the instru
47Phosphate HRPHOSPHATE HIGH RANGESPECIFICATIONSRange 0.0 to 30.0 mg/LResolution 0.1 mg/LAccuracy ±1 mg/L ±4% of reading at 25 °CTypical EMC Dev. ±0.1
48Phosphate HR• Add the content of one packet of HI 93717B-0 PhosphateHR Reagent B to the cuvette. Replace the cap and shakegently until completely di
49Phosphate LRPHOSPHATE LOW RANGESPECIFICATIONSRange 0.00 to 2.50 mg/LResolution 0.01 mg/LAccuracy ±0.04 mg/L ±4% of reading at 25 °CTypical EMC Dev.
5Where:-log I/Io= Absorbance (A)Io= intensity of incident light beamI = intensity of light beam after absorptionελ= molar extinction coefficient at wa
50Phosphate LR• Reinsert the cuvette into the instrument.• Press Timer and the display will show the countdown prior to the measurement or, alternativ
51PhosphorusPHOSPHORUSSPECIFICATIONSRange 0.0 to 15.0 mg/LResolution 0.1 mg/LAccuracy ±0.3 mg/L ±4% of reading at 25 °CTypical EMC Dev. ±0.2 mg/LLight
52Phosphorus• Add the content of one packet of HI 93706B-0 PhosphorusReagent B (Amino Acid) to the cuvette. Replace the capand shake gently until comp
53SilicaSILICASPECIFICATIONSRange 0.00 to 2.00 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of reading at 25 °CTypical EMC ±0.01 mg/LDeviationLight
54Silica• Place the cuvette into the holder and close the lid.• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready f
55• Press or to return to the measurement screen.INTERFERENCESInterference may be caused by:Phosphate above 60 mg/LPhosphate above 75 mg/LSulfide
56SilverSILVERSPECIFICATIONSRange 0.000 to 1.000 mg/LResolution 0.005 mg/LAccuracy ±0.005 mg/L ±10% of reading at 25 °CTypical EMC ± 0.001 mg/LDeviati
57Silver• Add exactly 1.0 mL of HI 93737B-0 Buffer reagent to the secondbeaker (the sample) and swirl gently to mix. Press Timer and thedisplay will s
58Silver• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.• Fill a second cuvette up to the mark
59ZincZINCSPECIFICATIONSRange 0.00 to 3.00 mg/LResolution 0.01 mg/LAccuracy ±0.03 mg/L ±3% of reading at 25 °CTypical EMC ±0.01 mg/LDeviationLight Sou
6INSTRUMENT DESCRIPTION1) Open Cuvette Lid2) Indexing mark3) Cuvette point4) Liquid Crystal Display (LCD)5) Splash proof keypad6) ON/OFF power switch7
60Zinc• Place the cap and insert the cuvette into the instrumentand close the lid.• Press the Zero key. The display will show “-0.0-” when the meter i
61INTERFERENCESInterference may be caused by:Aluminum above 6 mg/LCadmium above 0.5 mg/LCopper above 5 mg/LIron above 7 mg/LManganese above 5 mg/LNick
62The instrument shows clear warning messages when erroneous conditions appear and when measured values areoutside the expected range. These messages
63The analyzed data can be managed using Hanna’s product HI92000, Windows® Compatible Software.DATA MANAGEMENTWindows® is registered Trademark of &quo
64Description Range MethodAmmonia MR 0.00 to 10.00 mg/L NesslerAmmonia LR 0.00 to 3.00 mg/L NesslerChlorine, Free 0.00 to 2.50 mg/L DPDChlorine, Total
65ACCESSORIESREAGENT SETSHI 93700-01 100 ammonia LR testsHI 93700-03 300 ammonia LR testsHI 93701-01 100 free chlorine tests (powder)HI 93701-03 300 f
66OTHER ACCESSORIESHI 731318 cloth for wiping cuvettes (4 pcs)HI 731321 glass cuvettes (4 pcs)HI 731325W new cap for cuvette (4 pcs)HI 740034 cap for
67WARRANTYAll Hanna Instruments meters are warranted for two years against defects in workmanship and materialswhen used for its intended purpose and
68Printed in EUROPE(ROMANIA)Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USATechnical Support for CustomersTe
7KEYPAD DESCRIPTIONThe keypad contains 8 direct keys and 3 functional keys with the following functions:Press to perform the function displayed above
8The instructions listed below should be carefully followed during testing to ensure most accurate results.• Color or suspended matter in large amount
9• Proper use of the powder reagent packet:(a) use scissors to open the powder packet;(b) push the edges of the packet to form a spout;(c) pour out th