ERROR MESSAGES
The meter has lost its configuration. Contact your dealer or the nearest
Hanna Customer Service Center.
“Light high”: there is too much light to perform a measurement.
Please check the preparation of the zero cuvet.
“Light low”: there is not enough light to perform a measurement.
Contact your dealer or the nearest Hanna Customer Service Center.
“No Light”: the instrument cannot adjust the light level. Contact your
dealer or the nearest Hanna Customer Service Center.
a) on zero reading:a) on zero reading:
a) on zero reading:a) on zero reading:
a) on zero reading:
A flashing value of the maximum concentration indicates an over
range condition. The concentration of the sample is beyond the
programmed range: dilute the sample and measure again.
The sample absorbs less light than the zero reference. Check the pro-
cedure and make sure you use the same cuvet for reference (zero)
and measurement.
“Inverted”: the sample and the zero cuvet are inverted.
b) on sample reading:b) on sample reading:
b) on sample reading:b) on sample reading:
b) on sample reading:
Indicates that the batteries are dead and must be replaced. After this
message appears, the instrument is switched off. Change the batteries
and restart the meter.
19
After a few seconds the display will show “-0.0-”.
The meter is now zeroed and ready for measurement.
Remove the cuvet from the instrument.
Insert the sample (cuvet #2) into the holder and
close the lid.
#2
(Sample)
• Press READ/TIMER and “----” will blink during
measurement.
The instrument directly displays the color hue on the
Liquid Crystal Display.
Fill an empty cuvet with deionized water and replace
the cap. This is the
zerozero
zerozero
zero.
#1
(Zero)
Insert the zero cuvet into the holder and close the lid.
Press ZERO and “----” will blink on the display.
INTERFERENCES
Suspended matter should be removed by centrifugation or prior filtration. Use an adequate filter
material that does not absorb color.