Setup/Installation
This screen appears when the stirrer is
not working properly. Check the stir bar
and beaker content. Press RESTART to
try again.
This error message appears when the
pump is not working properly. Check the
tubing, valve and syringe. Press RESTART
to try again.
1. Remove the electrode protective
cap.
2. If the bulb is dry, soak the electrode
in HI 70300 Storage Solution for at
least one hour.
3. Preparation of reference electrode:
4. Unwrap Paralm™ seal found over
ceramic junction on inner stern of
the reference electrode and discard.
This is only used for shipping.
5. Rinse inner stern with deionized
water, making certain to wet o-ring
found on the inner stern.
6. Reassemble reference electrode by
gently pushing the inner assembly
into the outer body (from the
bottom), slide spring and cap down
the cable and over the top of the
inner stern. Screw the cap into
place.
7. Unscrew and remove the ll hole
Electrode
Preparation
Procedure
Titration
Messages