1Instruction ManualHI 84431TOTAL ALKALINITYMINITITRATOR & pH METERfor Water Analysiswww.hannainst.com

Pagina 2 - Dear Customer

10ESC - used to leave the current screen and to return either to the previous screen or to themain screen. When pressed while modifying a parameter wi

Pagina 3 - TABLE OF CONTENTS

11PERISTALTIC PUMPPeristaltic pumps are self priming. Liquid never contacts the pump components. The titrant tubingis pressed along the rotating rolle

Pagina 4 - GENERAL DESCRIPTION

12TITRATOR STARTUPThis is a general outline of the steps required to make a titration. The following sections expandupon each section.• Place the inst

Pagina 5

13From your knowledge of expected concentrations, use the table below to determine whichsettings and solutions are appropriate:Use Total LR or Strong

Pagina 6

14Use the ARROW keys in order to increase/decreasethe value.Press Accept to confirm or ESC to return to thesetup menu without saving the new value.pH

Pagina 7 - SPECIFICATIONS

15Press / to highlight the value to be modified(year, month, day, hour, minute or second). Usethe ARROW keys to change the value.Press Accept to

Pagina 8 - PRINCIPLE OF OPERATION

16Press the corresponding function key to change thelanguage.If the new selected language cannot be loaded,the previously selected language will be re

Pagina 9

17Press Select to view the firmware version, languageversion, mV factory calibration date and time andtemperature factory calibration date and time.Pr

Pagina 10

18ELECTRODE PREPARATIONPREPARATION PROCEDURERemove the electrode protective cap.DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This is normal wit

Pagina 11

19ELECTRODE CALIBRATION PROCEDUREIt is recommended to calibrate the instrument frequently, especially if high accuracy is required.The pH electrode sh

Pagina 12 - TITRATOR STARTUP

2Dear Customer,Thank you for choosing a Hanna product. This manual will provide you with the necessaryinformation for the correct use of the instrumen

Pagina 13 - SETUP CONFIGURATION MENU

20• The (unstable measurement) symbol will be shown on the display until the reading becomes stable.• When the reading is stable and close to the sele

Pagina 14

21• After the second calibration point has been confirmed, press ESC to exit without performingthe third calibration point.Point 3 calibration• Remove

Pagina 15

22• If the measured offset isn’t within the preset limits the meter will display the message“Buffer Contaminated” alternatively with ”Electrode Dirty/

Pagina 16

23The temperature has an effect on pH. The calibration buffer solutions are affected by temperaturechanges to a lesser degree than normal solutions. D

Pagina 17

24PURGEPurging should be performed:• whenever the tube of the peristaltic pump is replaced;• whenever the titrant is changed or a new bottle is used;•

Pagina 18 - ELECTRODE PREPARATION

25- a pump calibration hasn’t been performed or more than 3 days have passed since the lastpump calibration.• Press CAL to access the titration calibr

Pagina 19

26Verify: The electrode has been calibrated in 8.30 pH buffer.• Sample prepar.ation: Add a precisely measured amount ofHI 84431-55 Pump Calibration S

Pagina 20

27• Press Pump.The next screen will be displayed.• Press Start.• After the pump calibration is started, on the upper right side of the display two ani

Pagina 21

28• After the pump calibration is complete a confirmation message is displayed for a few seconds,then the instrument will return to the titrator calib

Pagina 22

29Verify: The instrument has been calibrated (pH and pump) before performing a lot oftitrations. An electrode calibration in at least one point (8.30

Pagina 23

3All rights are reserved. Reproduction in whole or in part is prohibited without the written consentof the copyright owner, Hanna Instruments Inc., Wo

Pagina 24 - PUMP TUBE INSTALLATION

30• After the titration is started on the upper right side of the display two animations will beshown in order to indicate that the pump and the stirr

Pagina 25

31VIEW/DELETE LOGGED SAMPLESPress MENU key while in Titrator main screen.Press Recall to access the titrator logged data.The instrument will display

Pagina 26 - PUMP CALIBRATION PROCEDURE

32Press Delete to enter delete record screen.Press Delete All to enter delete all records screen.Press More to view additional information of the mea

Pagina 27

33GLP contains a set of information regarding electrode calibration. The following items areincluded in electrode GLP: the time and date of the last c

Pagina 28

34If a calibration hasn’t been performed the message “Not Calibrated” will be displayed.The pump GLP displays the Time and Date of the last pump calib

Pagina 29 - TITRATION PROCEDURE

35Press LOG to memorize the current reading. A message indicating the free log space will bedisplayed for a few seconds.In order to take pH measureme

Pagina 30

36Press Confirm to delete the record or ESC to return to the previous screen without deleting.Deleting a record will renumber the list of records.VIEW

Pagina 31

37If the pH log is empty the message “No Records!” will be displayed.pH METER GLP INFORMATIONThe following items are included in electrode GLP: the ti

Pagina 32

38TEMPERATURE CALIBRATION PROCEDURE (for technical personnel only)All the instruments are factory calibrated for temperature.HANNA’s temperature probe

Pagina 33

39• Immerse the temperature probe in the second vessel as near as possible to the referencethermometer. Allow a few seconds for the probe to stabilize

Pagina 34

4Please examine this product carefully. Make sure that the instrument is not damaged. If anydamage occurred during shipment, please notify your Dealer

Pagina 35

40SMOTPMYSSMOTPMYSSMOTPMYSSMOTPMYSSMOTPMYSMELBORPMELBORPMELBORPMELBORPMELBORPNOITULOSNOITULOSNOITULOSNOITULOSNOITULOSevissecxe/esnopserwolS.tfird.edor

Pagina 36

41ELECTRODE CONDITIONING AND MAINTENANCEpH CLEANING PROCEDURE• GeneralSoak in Hanna HI 7061 or HI 8061 General Cleaning Solution forapproximately ½ ho

Pagina 37

42REAGENTSHI 84431-50 Titrant solution for low range (100mL)HI 84431-51 Titrant solution for high range (100mL)HI 84431-55 Pump calibration solution (

Pagina 38

43HI 84431 is warranteed for two years against defects in workmanship and materials when used for itsintended purpose and maintained according to the

Pagina 39

44Printed in Romania, EU MAN84431 08/09Hanna Instruments Inc.Highland Industrial Park584 Park East DriveWoonsocket, RI 02895 USALocal Sales and Custo

Pagina 40 - TROUBLESHOOTING GUIDE

5By simply pressing START key, the HI 84431 automatically starts pump operation and titrates thesample to the end point. The HI 84431 has a simple and

Pagina 41

6The results of a titration provide a means to classify the principle forms of alkalinity.Relationship of titrant volumes for alkalinity calculations:

Pagina 42 - ACCESSORIES

7SPECIFICATIONSTitrator Range Titratable alkalinity (low range)mg/L : 15.0 - 500.0 mg/L as CaCO3meq/L : 0.3 - 10.0 meq/L as CaCO3Titratable alkalinit

Pagina 43 - RECOMMENDATION FOR USERS

8PRINCIPLE OF OPERATIONWater alkalinity determination is performed by titration with acid reagent to pH end point.Alkalinity is expressed as:• Strong

Pagina 44 - Woonsocket, RI 02895 USA

9FUNCTIONAL AND PHYSICAL DESCRIPTIONOVERHEAD VIEWOVERHEAD VIEWOVERHEAD VIEWOVERHEAD VIEWOVERHEAD VIEW1. Titrant bottle2. Graphic LCD3. Keypad4. Dedica