25
24£
24
HI 8510E HI 8510T
Range 0.00 to 14.00 pH
Resolution 0.01 pH
Accuracy ±0.02 pH ±0.5%
Typical EMC Dev. ±0.1 pH / ±0.2 mA
Installation Category II
Input 10
12
Ohm 4 to 20 mA
Calibration Offset: ±2 pH with Δ0 trimmer
Slope: 80 to 110% with slope trimmer
Temperature Fixed or automatic with Pt100
Compensation from 0 to 100°C (32 to 212°F)
Recorder Output 0-20 mA or 4-20 mA (isolated)
Power Supply 115 or 230 Vac; 50/60 Hz
Environment -10 to 50°C (14 to 122°F);
RH max 95% non condensing
Panel Cutout 141 x 69 mm (5.6 x 2.7'')
Weight 1 kg (2.2 lb.)
SPECIFICATIONS
HI 8710E HI 8710T
Range 0.00 to 14.00 pH
Resolution 0.01 pH
Accuracy ±0.02 pH ±0.5%
Typical EMC Dev. ±0.1 pH / ±0.2 mA
Installation Category II
Input 10
12
Ohm 4 to 20 mA
Calibration Offset: ±2 pH with Δ0 trimmer
Slope: 80 to 110% with slope trimmer
Temperature Fixed or automatic with Pt100
Compensation from 0 to 100°C (32 to 212°F)
Relays 1 for setpoint and 1 for alarm,
max 2A, 240 V resistive load (isolated)
Recorder Output 0-20 mA or 4-20 mA (isolated)
Power Supply 115 or 230 Vac; 50/60 Hz
Environment -10 to 50°C (14 to 122°F);
RH max 95% non condensing
Panel Cutout 141 x 69 mm (5.6 x 2.7'')
Weight 1 kg (2.2 lb.)
HI 8512E HI 8512T
Range ±1000 mV
Resolution 1 mV
Accuracy ±5 mV ±0.5%
Typical EMC Dev. ±6 mV / ±0.2 mA
Installation Category II
Input 10
12
Ohm 4 to 20 mA
Calibration Slope: 90 to 110% with slope trimmer
Recorder Output 0-20 mA or 4-20 mA (isolated)
Power Supply 115 or 230 Vac; 50/60 Hz
Environment -10 to 50°C (14 to 122°F);
RH max 95% non condensing
Panel Cutout 141 x 69 mm (5.6 x 2.7'')
Weight 1 kg (2.2 lb.)
HI 8711E HI 8711T
Range 0.00 to 14.00 pH
Resolution 0.01 pH
Accuracy ±0.02 pH ±0.5%
Typical EMC Dev. ±0.1 pH / ±0.2 mA
Installation Category II
Input 10
12
Ohm 4 to 20 mA
Calibration Offset: ±2 pH with Δ0 trimmer
Slope: 80 to 110% with slope trimmer
Temperature Fixed or automatic with Pt100
Compensation from 0 to 100°C (32 to 212°F)
Relays 2 for setpoint and 1 for alarm,
max 2A, 240 V resistive load (isolated)
Recorder Output 0-20 mA or 4-20 mA (isolated)
Power Supply 115 or 230 Vac; 50/60 Hz
Environment -10 to 50°C (14 to 122°F);
RH max 95% non condensing
Panel Cutout 141 x 69 mm (5.6 x 2.7'')
Weight 1 kg (2.2 lb.)