HI8417 - HI8519
HI8520 - HI8521
Microprocessor
Bench-top
pH Meters
Instruction Manual
http://www.hannainst.com
These Instruments are in
Compliance with the CE Directives
HI 8417
°C
pH
mV
pH
OFF
ON
CON
CAL
°C
°C
d2
t2
d1
t1
pH
pH
6,88
7,02
4,00
4,00
9,22
10,05
pH
pH
pH
pH
BUFFER VALUES 20°C
NBS
STD
NBS
STD
NBS
STD
d3
t3
Vista de pagina 0
1 2 3 4 5 6 ... 39 40

Indice de contenidos

Pagina 1 - Bench-top

HI8417 - HI8519HI8520 - HI8521MicroprocessorBench-toppH MetersInstruction Manualhttp://www.hannainst.comThese Instruments are inCompliance with the CE

Pagina 2 - Company since 1992

10RANGEOFFTemperature Buffer Values (pH)°C°F4.01 7.01 10.01010203040325068861044.014.004.004.024.047.137.077.037.006.9810.3210.1810.069.969.88CONERROR

Pagina 3 - PRELIMINARY EXAMINATION

11off. Press the RESET button and restart the entire operation.Power connectionPlug the 12VDC adapter (HI710005 or HI710006) into the socket onthe bac

Pagina 4

12COMBINEDpH or ORPELECTRODESREFERENCEELECTRODERESETTEMPERATUREPROBE12345DC POWER12V to 20V+RANGEOFFTemperature Buffer Values (pH)°C°F4.01 7.01 10.010

Pagina 5 - SPECIFICATIONS OF HI8417

13instrument's power supply is disconnected before the meter is switchedoff. Press the RESET button and restart the entire operation.Power connec

Pagina 6 - THE REAR PANEL OF HI8417

14OPERATIONAL GUIDETAKING PH MEASUREMENTSConnect the instrument to the 12VDC adapter (HI710005 orHI710006). Plug the adapter into the mains.Make sure

Pagina 7

154 cm(1½")°CpH• In HI8417 the temperature reading is also displayed on the leftLCD.Note: the electrode should be submerged approximately 4 cm (1

Pagina 8 - SPECIFICATIONS OF HI8519

16• When the temperature is set manually,"°C" symbol blinks. With the tempera-ture probe connected, the "°C" symbolis displayed co

Pagina 9

17For HI8417 and HI8521• Connect the ORP electrode to the meter and submerge the tip (4cm / 1½") into the sample to be tested.Note: ORP measureme

Pagina 10 - THE REAR PANEL OF HI8520

18IMPORTANTThe instrument's pH range should be re-calibrated:• When the meter is new.• Whenever the pH electrode is replaced.• At least once a mo

Pagina 11

19Procedure:• Switch the instrument on after connecting thepH electrode and the temperature probe.• Immerse the pH electrode and the ATC probeinto a n

Pagina 12 - THE REAR PANEL OF HI8521

2Dear Customer,Thank you for choosing a Hanna Instruments Product.Please read this instruction manual carefully before using the instru-ment.This manu

Pagina 13

20• The meter automatically recognizes the buffer value and promptsfor the second at "d2" or "t2" (= pH 4.01, acid) or "d3&qu

Pagina 14 - TAKING PH MEASUREMENTS

21Procedure:• Switch the instrument on after connecting the pHelectrode.• Rinse and immerse the pH electrode in pH 7.01buffer and shake briefly. Wait

Pagina 15 - HI8520 AND HI8521

22• Press the CAL key. The temperature compensatedbuffer value will be displayed e.g. if the buffertemperature is at 25°C, the display will showpH7.01

Pagina 16 - TAKING ORP MEASUREMENTS

23PH VALUES AT VARIOUS TEMPERATURETemperature has an effect on pH. The calibration buffer solutions areeffected by temperature changes to a lesser deg

Pagina 17 - AFTER MEASUREMENTS

24Possible Solutions:Replace the probe.Replace the electrode andrepeat the pH calibration.Leave the electrode in astorage solution after clean-ing the

Pagina 18 - PH CALIBRATION FOR HI8417

25ReferenceFilling HoleReferenceFilling HoleSensitiveWireReferenceWireReferenceJunctionGlassBulbGlassBulbReferenceJunctionReferenceJunctionReferenceJu

Pagina 19

26For refillable electrodes:If the filling solution (electrolyte) is more than 2½ cm (1") below thefill hole, add HI 7082 or HI 8082 3.5M KCl Ele

Pagina 20 - PH CALIBRATION FOR HI8519

27CLEANING PROCEDUREGeneral Soak in Hanna HI 7061 or HI 8061 General Clean-ing Solution for approximately ½ hour.Removal of films, dirt or deposits on

Pagina 21 - TEMPERATURE °C

28TEMPERATURE-RESISTANCECORRELATIONThe resistance of glass electrodes partially depends on the tempera-ture. The lower the temperature, the higher the

Pagina 22

29Alkaline Error1.0 Mol L-1 Na+0.1 Mol L-1 Na+Sodium Ion Correction for the Glass at 20-25°CConcentration pH Error13.0013.5014.0012.5013.0013.5014.0

Pagina 23 - (HI8417, HI8520, HI8521 ONLY)

3HI8417, HI8519, HI8520 and HI8521 Hanna Instruments bench-top pH meters are designed for simplicity of use.They feature: large liquid crystal display

Pagina 24 - TROUBLESHOOTING

30ACCESSORIESpH CALIBRATION SOLUTIONSHI 70004P pH 4.01 Buffer Sachets, 20mL, 25 pcsHI 70007P pH 7.01 Buffer Sachets, 20mL, 25 pcsHI 70010P pH 10.01 Bu

Pagina 25 - PREPARATION

31HI 7082 3.5M KCl Electrolyte, 4x50 mL, for double junctionelectrodesREFILL ELECTROLYTE SOLUTIONS IN FDA APPROVEDBOTTLESHI 8071 3.5M KCl + AgCl Elect

Pagina 26 - PERIODIC MAINTENANCE

32120 mm4.7"12 mm0.5"5 mm0.2"3 mm0.12"3.0 mm DIA0.12"HI 1083120 mm4.7"12 mm0.5"9.5mm DIA0.37""S" VER

Pagina 27 - CLEANING PROCEDURE

3375 mm2.95"6 mm0.25""S"VERSIONHI 2031120 mm4.7"12 mm0.5""S" VERSIONHI 1332120 mm4.7"12 mm0.5"FC 100

Pagina 28 - CORRELATION

34120 mm4.7"12 mm0.5"FC 210120 mm4.7"12 mm0.5"9.5mm DIA0.37"FC 220FC 210BGlass-body, double junction, conic, Viscolene, non-r

Pagina 29

35150 mm5.9"12 mm0.5""S" VERSIONHI 3131120 mm4.7"12 mm0.5"HI 3230"S" VERSION120 mm4.7"12 mm0.5"HI 44

Pagina 30 - ACCESSORIES

36HI7855/10 Extension cable 10m (33') longHI7855/15 Extension cable 15m (49.5') longPower Units:HI710005 Voltage adapter from 115 VAC to 12

Pagina 31 - ORP PRETREATMENT SOLUTIONS

37ELECTRODE APPLICATIONREFERENCE GUIDEApplication Electrodes1. Aquarium HI 1332B, HI 1911B2. Bath-water HI 1910B, HI 1130B3. Beer HI 1131B4. Bread HI

Pagina 32 - "S" VERSION

38WARRANTYAll Hanna Instruments meters are guaranteed for two years againstdefects in workmanship and materials when used for their intendedpurpose an

Pagina 33 - 0.25"

39CE DECLARATION OF CONFORMITYRecommendations for UsersBefore using these products, make sure that they are entirely suitable for the environment in w

Pagina 34 - 0.5"

4HI8520 pH/°C meter automatically recognizes 3 standard calibrationbuffer values (4.01, 7.01 and 10.01 pH).HI8520 comes supplied with HI1332B plastic-

Pagina 35 - 4.7"

MANPHBNR3 05/98http://www.hannainst.comPRINTED IN PORTUGAL

Pagina 36 - OTHER ACCESSORIES

5HI 8417°CpHmVpHOFFONCONCAL°C°Cd2t2d1t1pHpH6,887,024,004,009,2210,05pHpHpHpHBUFFER VALUES 20°CNBSSTDNBSSTDNBSSTDd3t3THE FRONT PANEL OF HI8417ON / OFF

Pagina 37 - REFERENCE GUIDE

6COMBINEDpH or ORPELECTRODESREFERENCEELECTRODERESETRECORDER0-2 VOLTTEMPERATUREPROBE123456DC POWER12V to 20V+THE REAR PANEL OF HI84171. BNC socket for

Pagina 38 - WARRANTY

7instrument's power supply is disconnected before the meter is switchedoff. Press the RESET button and restart the entire operation.Power connect

Pagina 39 - CE DECLARATION OF CONFORMITY

8TemperatureBuffer Values (pH)°C°F4.01 7.0110.01010203040325068861044.014.004.004.024.047.137.077.037.006.9810.3210.1810.069.969.88pHmV°COFFSETTEMPERA

Pagina 40 - PRINTED IN PORTUGAL

9Electrode and probe connectionsFor combination pH or ORP electrodes (with internal reference) plugthe electrode's BNC to the socket provided (#1