Temperature has an effect on the pH. The calibration buffer solutions
are affected by temperature changes to a lesser degree than normal
solutions.
Please refer to the following chart to perform the pH calibration:
For instance, if the buffer temperature is 25°C (77°F), calibrate to
read on the display pH 4.01 or 7.01 or 10.01.
If the buffer temperature is 20°C, calibrate to read on the display
pH 4.00 or 7.03 or 10.06.
If the buffer temperature is 50°C, calibrate to read on the display
pH 4.06 or 6.98 or 9.82.