Hanna-instruments HI 901 Manual de usuario Pagina 135

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 148
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 134
A3-3
APPENDIX 3
A3 MULTIPLE EQUIVALENCE POINTS (HI 902 only)
A3.1 Applicability Domain
In case of:
determining the components concentration from a synthetic mixture (e.g. Mixture of
HCl, CH
3
COOH, NH
4
Cl).
determining the concentration of a polyprotic acid in its titrable ionization stages
(e.g. H
3
PO
4
, two equivalence points in aquogenos medium).
A3.2 Perform a Multi-Equivalence Point Titration
With the information provided above, set up the method parameters as follows:
 Set the end point mode as equivalence end point (mV or pH).
 Set the number of equivalence points (up to 5 points) to be found.
 Select the end point determination (like in standard titration).
 Set the rest of method parameters and save the modification.
Vista de pagina 134
1 2 ... 130 131 132 133 134 135 136 137 138 139 140 ... 147 148

Comentarios a estos manuales

Sin comentarios