* 1 FTU = 1 NTU
Even though HI93703-11 covers a very
wide range of turbidity values, for very
accurate measurements of samples ex-
ceeding 40FTU*, Standard Methods re-
quire dilution. In such cases, the correct
amount of HI93703-0 or turbidity-free
water to be added to the sample can be
calculated as follows:
Vos = 3000 / T
where: Vos = volume of sample (mL) to be
combined with HI 93703-0 to
obtain the final volume of 100 mL.
T = HI 93703 reading (exceeding
40FTU*)
E.g.: HI 93703 reading = 200 FTU *
3000 / 200 = 15 mL (Vos)
15 mL (Vos)+85 mL (HI 93703-0) = 100 mL
At this point, take a sample of this solution
and measure its turbidity.
The correct turbidity value of the original
sample will be:
T
n
x 100 mL / Vos = T
a
where: T
n
= new HI 93703 reading
T
a
= actual turbidity value of the original
sample
E.g. If T
n
= 27 FTU *
T
a
= 27 FTU *x100 mL / 15 mL = 180 FTU *
NOTE: Any sample taken above 1000 FTU will
show out of range by blinking "1000".
15
* 1 FTU = 1 NTU
hours (the sample needs to be kept at room
temperature prior to the analysis).
To obtain a representative sample, gently,
but thoroughly, mix it before samples are
taken. Do not shake (to prevent air bubbles)
and do not let the sample settle.
It is recommended to calibrate the meter with
the supplied HI93703-10 @10 FTU* stan-
dard at least once a month or more frequently
for greater accuracy.
Before inserting vials into the instrument,
wipe them with HI93703-70 or a soft, lint-
free tissue. Handle vials so that no finger-
prints get on the areas where light passes
(approx. the bottom 2 cm/1 inch).
If you experience any problems in taking mea-
surements, contact your dealer or the nearest
Hanna Customer Service Center.
SOURCES OF INTERFERENCE
Presence of floating debris and coarse sedi-
ments which settle out rapidly will give false
readings.
The infrared light source used for HI 93703-
11 turbidity meter, complies with the ISO
7027 International Standard and can effec-
tively minimize any errors due to colored
dissolved substances. This effect, called
"true color", is a common interference for
most commercially available instruments
operating in the range of visible light.
Air bubbles and the effect of vibrations that
disturb the surface of the sample may give
false results and should be avoided.
Dirty glassware could also affect readings
along with scratched or dented vials.