Hanna-instruments HI 93710 Manual de usuario Pagina 35

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 52
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 34
35
Reinsert the cuvet into the instrument.
Wait for approximately 30 seconds and then press READ. The
display will show “SIP” during measurement.
The instrument directly displays concentration in mg/L of free
chlorine on the Liquid Crystal Display.
Note: Free and total chlorine have to be measured separately with
fresh unreacted samples following the above procedure if both
values are requested.
INTERFERENCES
Interference may be caused by:
Bromine
Iodine
Fluorine
Ozone
Oxidized manganese and Chromium
In case of water with hardness greater than 500 mg/L CaCO
3
, shake
the sample for approximately 1 minute after adding the reagent.
In case of water with alkalinity greater than 250 mg/L CaCO
3
or
acidity greater than 150 mg/L CaCO
3
, the color of the sample could
disappear or develop only partially. To resolve this, neutralize the
sample with diluted HCl or NaOH.
Free Chlorine
Vista de pagina 34
1 2 ... 30 31 32 33 34 35 36 37 38 39 40 ... 51 52

Comentarios a estos manuales

Sin comentarios