Liquid reagent procedure
Add 3 drops of HI 93701A-T DPD1 indicator, 3 drops of
HI 93701B-T DPD1 buffer and 1 drop of HI 93701-C to another
cuvet. Shake gently before adding 10 mL of unreacted sample.
Replace the cap and shake gently again.
Reinsert the cuvet into the instrument.
Wait for approximately 30 seconds and then press READ. The
display will show "SIP" during measurement.
The instrument directly displays concentration in mg/L of free or
total chlorine on the Liquid Crystal Display.
Note: Free and total chlorine have to be measured separately with
fresh unreacted samples following the above procedure if both
values are requested.
INTERFERENCES
Interference may be caused by:
Bromine, Iodine, Fluorine, Ozone, Oxidized manganese and Chro-
mium.
In case of water with hardness greater than 500 mg/L CaCO
3
, shake
the sample for approximately 1 minute after adding the reagent.
Alkalinity above 250 mg/L or acidity above 150 mg/L will not
reliably develop the full amount of color or it may rapidly fade. To
resolve this, neutralize the sample with diluted HCl or NaOH.
Total Chlorine
x 3 x 3
x 1