1www.hannainst.comThis Instrument is inCompliance with the CE DirectivesInstruction ManualHI 95747CCopper Low Range ISM

Pagina 2 - TABLE OF CONTENTS

10The flashing digits indicate that:1) the month can be set2) the day can be set“Storage”: this prompt appears for 1 second at the end of thecalibrati

Pagina 3 - PRELIMINARY EXAMINATION

11The instructions listed below should be carefully followed during testing to ensure best accuracy.• Color or suspended matter in large amountsmay ca

Pagina 4 - ABBREVIATIONS

12• It is important that the sample does not containany debris. This would corrupt the reading.• In order to avoid reagent leaking and to obtainmore a

Pagina 5 - PRECISION AND ACCURACY

13MEASUREMENT PROCEDURE• Turn the meter on by pressing ON/OFF.• When the LCD displays “747”, it is ready.• Fill the cuvet with 10 mL of unreacted samp

Pagina 6 - PRINCIPLE OF OPERATION

14• Hold READ/TIMER for three seconds. Thedisplay will show the countdown prior tomeasurement.Alternatively, wait for 45 seconds and justpress READ/TI

Pagina 7 - FUNCTIONAL DESCRIPTION

15• Turn the meter on by pressing ON/OFF.When the LCD displays “747”, it is ready.• Place the CAL CHECKTM Standard Cuvet A intothe holder and ensure t

Pagina 8 - Calibration mode:

16CALIBRATION PROCEDURE• Turn the meter on by pressing ON/OFF.When the LCD displays “747”, it is ready.• Enter the calibration mode by holding CALCHEC

Pagina 9 - GUIDE TO DISPLAY CODES

17• Press READ/TIMER and “SIP” will blink duringmeasurement.• The instrument will show for three seconds theCAL CHECKTM standard value.Then the date o

Pagina 10

18REAGENT SETSHI 95747-01 Reagents for 100 testsHI 95747-03 Reagents for 300 testsOTHER ACCESSORIESHI 95747-11 CAL CHECKTM Standard Cuvets (1 set)

Pagina 11 - (a) (b) (c)

19Recommendations for UsersBefore using these products, make sure that they areentirely suitable for your specific application and for theenvironment

Pagina 12

2TABLE OF CONTENTSPRELIMINARY EXAMINATION ... 3GENERAL DESC

Pagina 13 - MEASUREMENT PROCEDURE

20SALES AND TECHNICAL SERVICE CONTACTSMAN95747CR1 11/03Australia:Tel. (03) 9769.0666 • Fax (03) 9769.0699China:Tel. (10) 88570068 • Fax (10) 8857

Pagina 14

3PRELIMINARY EXAMINATIONPlease examine this product carefully. Make sure that the instrument is not damaged. If anydamage occurred during shipment, pl

Pagina 15 - VALIDATION PROCEDURE

4The HI 95747 is an auto diagnostic portable microprocessor meter that benefits from Hanna’syears of experience as a manufacturer of analytical instru

Pagina 16 - CALIBRATION PROCEDURE

5Range 0.000 to 1.500 mg/LResolution 0.001 mg/L for measurement0.01 mg/L for calibration and validationPrecision ±0.015 mg/L @ 0.750 mg/LTypical EMC D

Pagina 17

6Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation andmatter. When a light beam crosses a substance, some o

Pagina 18 - BATTERY REPLACEMENT

7A microprocessor controlled special tungsten lamp emits radiation which is first opticallyconditioned and beamed to the sample contained in the cuvet

Pagina 19 - CE DECLARATION OF CONFORMITY

8KEYPAD DESCRIPTION• ON/OFF: to turn the meter on and off.• ZERO: to zero the meter prior to measurement.• READ/TIMER: this is a bi-functional key, ju

Pagina 20 - MAN95747CR1 11/03

9This prompt appears for 1 second each time the instrument isturned on.This indicates that the meter is in a ready state and zeroing canbe performed.S