Hanna-instruments HI 96700 Manual de usuario Pagina 14

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 24
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 13
14
Alternatively, wait for 3 minutes and 30
seconds then just press READ/ /TIMER.
In all cases, the lamp, cuvette and detector
icons will appear on the display,
depending on the measurement phase.
At the end of measurement, the instrument
directly displays concentration in mg/L of
ammonia nitrogen (NH
3
-N) on the LCD.
To convert the reading to mg/L of
ammonia (NH3), multiply the display
by a factor of 1.214.
INTERFERENCES
Interference may be caused by:
Hardness above 1 g/L
Iron, Sulfide, Glycine, various aliphatic and aromatic amines
Organic chloramines
Acetone
Aldehydes
Alcohols.
Vista de pagina 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 ... 23 24

Comentarios a estos manuales

Sin comentarios