When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
The blinking “ZERO” indicates that the
instrument needs to be zeroed first.
Fill one cuvette with 10 mL of unreacted
sample, up to the mark.
Fill a second cuvette with 10 mL of
distilled water, up to the mark (this is
the blank).
Add 12 drops of the HI 93704-0 reagent
to each cuvette. Replace the caps and
swirl the solutions.
Place the blank into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
Press and hold READ/ /TIMER for three
seconds. The display will show the
countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.
Alternatively, wait for 12 minutes then
just press ZERO/CFM.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.
10 mL
sample
10 mL
blank
12
3
6
9
12’