Hanna-instruments HI 98170 Manual de usuario Pagina 13

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 52
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 12
13
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the
selected buffer, “CFM” tag blinks.
Press CFM to confirm calibration.
The calibrated value is then displayed on the
primary LCD and the secondary LCD will display
the second expected buffer value, together with
“CAL” and “Cal Point 2” tags.
After the first calibration point is confirmed, immerse the pH electrode and the
temperature probe approximately 4 cm (1½”) into the second buffer solution
and stir gently. The temperature probe should be close to the pH electrode.
If necessary, press the ARROW keys to select a different buffer value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected buffer, “CFM” tag blinks.
Press CFM to confirm calibration.
The calibrated value is then displayed on the
primary LCD and the secondary LCD will display
the third expected buffer value.
After the second calibration point is confirmed, immerse the pH electrode
and the temperature probe approximately 4 cm (1½”) into the third
buffer solution and stir gently. The temperature probe should be close to
the pH electrode.
If necessary, press the ARROW keys to select a different buffer value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected buffer, “CFM” tag blinks.
Press CFM to confirm calibration.
The calibrated value is then displayed on the
primary LCD and the secondary LCD will display
the fourth expected buffer value.
After the third calibration point is confirmed, immerse the pH electrode
and the temperature probe approximately 4 cm (1½”) into the fourth
buffer solution and stir gently. The temperature probe should be close to
the pH electrode.
If necessary, press the ARROW keys to select a different buffer value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected buffer, “CFM” tag blinks.
Press CFM to confirm calibration.
The calibrated value is then displayed on the
primary LCD and the secondary LCD will display
the fifth expected buffer value.
LOG CFM
LOG CFM
LOG CFM
LOG CFM
Vista de pagina 12
1 2 ... 8 9 10 11 12 13 14 15 16 17 18 ... 51 52

Comentarios a estos manuales

Sin comentarios