Actual Conductivity
Select the “Actual conductivity” option and
press <OK> to confirm.
Insert the custom value with the desired reso-
lution. Press <OK> to confirm.
Immerse the probe in the conductivity solution and wait for stable reading. The
probe sleeve must be inserted.
When the reading is stable, press <Confirm> to store the calibration data.
After confirmation, the following messages are displayed: “Storing data on
probe, please wait...”, “Updating GLP data, please wait...” and “Calibration
completed”.
Press <OK> to return to the “Calibration” menu.
To return to the main menu, press ESC repeatedly.
Salinity
Select the “Salinity” option and press <OK>.
Insert the salinity value of the custom calibra-
tion solution.
Immerse the probe in the solution and wait for
stable reading. The probe sleeve must be in-
serted.
• When the reading is stable, press <Confirm> to store the calibration data.
After confirmation, the following messages are displayed: “Storing data on
probe, please wait...”, “Updating GLP data, please wait...” and “Calibration
completed”.
Press <OK> to return to the “Calibration” menu.
To return to the main menu, press ESC repeatedly.
NotesNotes
NotesNotes
Notes These procedures calibrate the slope value. To calibrate the offset, set
the calibration point at 0 µS/cm and repeat the procedure.
If the temperature input is not within the acceptable range (0 to 50°C),
the message “Invalid temperature” is displayed.
If the conductivity input is not within
the acceptable range, the message “In-
valid input” is displayed.