After confirmation, the following messages are displayed: “Storing data on
probe, please wait...”, “Updating GLP data, please wait...” and “Calibration
completed”.
Press OK to return to the “Calibration” menu.
To return to the main menu, press ESC repeatedly.
To restore the factory calibration data, select the corresponding option in the
ORP calibration” menu and then press <OK>.
4.7 TEMPERATURE CALIBRATION
The meter is factory calibrated for temperature readings. If necessary, tempera-
ture calibration may be performed as explained below.
Select “Temperature” option from the “Calibration” menu and press <OK> to
enter the temperature calibration mode.
Insert the probe in a thermostatic bath.
Select the measure unit (°C, °F or K) and in-
sert the bath temperature value (measured with
a reference thermometer).
When the reading in stable, “Ready” appears
on the LCD.
Press <Confirm> to store the calibration point.
After confirmation, the following messages are
displayed: “Storing data on probe, please
wait...”, “Updating GLP data, please wait...”
and “Calibration completed”.
Press <OK> to return to the “Calibration” menu.
To return to the main menu, press ESC repeatedly.
NoteNote
NoteNote
Note The meter allows a maximum difference
of ±2°C between the current reading
and the set value. If this condition is
not satisfied, the display shows the
warning message “Max +/-2°C is al-
lowed”.