Alternatively, press LOG/CFM to skip the first calibration point.
Then the LCD will show the second calibration
point (15.0 FTU) and “CAL P.2”, while “READ”
is blinking.
Note: If you’re using different calibration
standards, change the displayed value by
pressing UP or DOWN keys until the display
shows the desired value.
Remove the <0.1 FTU standard cuvet and
place the 15.0 FTU standard cuvet into the
holder.
Close the lid and press READ . “----” and the
“LED and Read Status” indicator will blink
again while making the reading.
At the end of the reading, the third calibration
point (100 FTU) and “CAL P.3” is displayed. If
desired, the value can be changed using the
UP or DOWN keys.
Note: At this moment it is possible to exit
calibration by pressing CAL/RCL. The instrument
will memorize the two-point (<0.1 and 15.0
FTU) calibration data and will return to
measurement mode.
Remove the 15.0 FTU standard cuvet from the
meter and place the 100 FTU standard cuvet
into the holder.
Close the lid and press READ . “----” and the
“LED and Read Status” indicator will blink
again while making the reading.
<0.1
FTU
15.0
FTU
15.0
FTU
100
FTU