13
At the end of the reading, the fourth calibration
point (800 FTU) and “CAL P.4” is displayed. If
desired, the value can be changed using the
UP or DOWN keys.
Note: At this moment it is possible to exit calibration by pressing CAL/RCL. The instrument will
memorize the three-point (<0.1, 15.0 and 100 FTU) calibration data and will return to
measurement mode.
or
Remove the 100 FTU standard cuvet from the
meter and insert the 800 FTU standard cuvet
into the holder.
Close the lid and press READ . “----” and the
“LED and Read Status” indicator will blink again
while making the reading.
At the end of the measurement, the four-point
calibration is completed and the instrument
returns automatically to measurement mode.
OUT CAL RANGE FUNCTION
The instrument has an Out Cal Range function to prevent
taking measurements in a range where the calibration
do not assure the best results. The range where the
calibration assures correct measurements is up to 40 FTU
for a two-point calibration and up to 150% of the third
point value for a three-point calibration.
The display will show a blinking “CAL” tag each
time the measurements are taken outside the
calibration range.
CALIBRATION ERRORS
If the read value during calibration is too far
from the set value, the instrument will show
-LO- or -HI- error.
100
FTU
800
FTU