1312
After confirmation, the selected param-
eter is displayed. The user can scroll
through the parameters by pressing CFM.
In order to directly set another param-
eter, press SETUP again and enter the
code or scroll to it by pressing CFM.
The following table lists the setup codes along with the descrip-
tion of the specific setup items, their valid values and whether
password is required to view that item (“PW” column):
Code Valid Values Default PW
00 Factory ID 0 to 9999 0000 no
01 Process ID 0 to 9999 (models w/o RS485) 0000 no
0 to 99 (models with RS485) 00 no
02 Control enable/disable 0: C.M. disabled 0 no
1: C.M. enabled
11 Relay 1 mode 0: disabled 0 no
(M1) 1: ON-OFF high setpoint
2: ON-OFF low setpoint
3: PID, high setpoint
4: PID, low setpoint
for pH 5025YZ and mV 6025YZ only:
0: disabled
1: PID, high setpoint
2: PID, low setpoint
12 Relay 1 setpoint 0.00 to 14.00 pH 8.00 pH no
(S1) -2000 to 2000 mV 500 mV
13 Relay 1 hysteresis 0.00 to 14.00 pH 1 pH no
(H1) 0 to 4000 mV 50 mV
14 Relay 1 deviation 0.50 to 14.00 pH 1 pH no
(D1) 25 to 4000 mV 50 mV
15 Relay 1 reset time 0.1 to 999.9 minutes 999.9 mins no
16 Relay 1 rate time 0.0 to 999.9 minutes 0.0 mins no
21* Relay 2 mode (M2) same as relay 1 0 no
* Available only in models with two relays
Then confirm the displayed digit with
and move to the next one.
When the whole password has been
inserted, press CFM to confirm it.
Note The default password is set at “0000”.
The LCD will display “SET” on the
upper part and “c.00” on the lower,
allowing the user to edit setup pa-
rameters (see table below).
Enter the code of the parameter you want to set, using the
arrow keys as per the password procedure above (e.g. 41).
Confirm the code by pressing CFM and the default or the
previously memorized value will be displayed with the first
digit blinking.
Note When the password is not inserted or a wrong password is
confirmed, the display will only show the previously memo-
rized value, without blinking (read only mode). In this case,
the value cannot be set. Press LCD and start again.
Enter the desired value using the arrow keys and then press
CFM.