62 63
• Add 12 drops of the HI 93704-0 reagent to
each cuvette. Replace the caps and shake
gently to mix (about 30 seconds).
• Place the blank (#1) into the holder and close the lid.
• Press Timer and the display will show the countdown prior to zeroing the blank. The display will show
“-0.0-” when the meter is zeroed and ready for measurement.
• Remove the blank.
• Insert the cuvette with the reacted sample (# 2) into the
instrument and close the lid.
• Press Read to start the reading. The instrument displays concentration in
µµ
µµ
µg/L of hydrazine.
Hydrazine
INTERFERENCES
Interference may be caused by:
Highly colored samples
Highly turbid samples
Aromatic amines
Hydrazine
HYDRAZINE
SPECIFICATIONS
Range 0 to 400 µg/L
Resolution 1 µg/L
Accuracy ±4% of full scale reading at 25 °C
Typical EMC ±2 µg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter @ 420 nm
Method Adaptation of the
ASTM Manual of Water and Environmental Technology, method
D1385-88,
p-Dimethylaminobenzaldehyde method. The reaction between hydrazine
and the liquid reagent causes a yellow tint in the sample.
REQUIRED REAGENT
Code Description Quantity
HI 93704-0 Liquid Reagent 24 drops
REAGENT SETS
HI 93704-01 Reagents for 100 tests
HI 93704-03 Reagents for 300 tests
For other accessories see page 124.
MEASUREMENT PROCEDURE
• Select the
Hydrazine
method using the procedure described in
the
Method Selection
section (see page 12).
• Fill one cuvette up to the mark with 10 mL of distilled water.
• Place the cap, insert the cuvette # 1 into the holder and
close the lid.
• Press the Blank key to start adjusting the light level. The display will show “Blank done” when the
meter is ready to take a zero measurement.
• Fill a second cuvette with 10 mL of unreacted sample (up to
the mark) and replace the cap.
# 1 # 2
# 1
# 2
10 mL
# 1
10 mL
# 2
# 1