Magnesium
SPECIFICATIONS
Range 0 to 150 mg/L
Resolution 5 mg/L
Accuracy ±5 mg/L ±3% of reading at 25 °C
Light Source Tungsten lamp with narrow band interference filter @ 466 nm
Method Adaptation of the Calmagite method
.
REQUIRED REAGENTS
Code Description Quantity
HI 93752A-0 Mg Magnesium Buffer Reagent 1 mL
HI 93752B-0 Mg Magnesium Indicator Reagent 9 mL
REAGENT SETS
HI 937520-01 Reagents for 50 tests
HI 937520-03 Reagents for 150 tests
For other accessories see page 124.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the COLORED OR TURBID SAMPLES
procedure on page 17.
Select the
Magnesium
method using the procedure described in the
Method Selection
section (see page 12).
Using one 1 mL syringe add exactly 1.00 mL of HI 93752A-0
Mg Buffer reagent to the cuvette and use the pipette to fill the
cuvette up to the 10 mL mark with the HI 93752B-0 Mg
Indicator reagent.
Replace the cap and invert several times to mix.
Place the cuvette into the holder and close the lid.
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
MAGNESIUM
Iron LR
INTERFERENCES
Interference may be caused by:
Cadmium above 4.0 mg/L
Chromium
3+
above 0.25 mg/L
Chromium
6+
above 1.2 mg/L
Cobalt above 0.05 mg/L
Copper above 0.6 mg/L
Cyanide above 2.8 mg/L
Manganese above 50.0 mg/L
Mercury above 0.4 mg/L
Molybdenum above 4.0 mg/L
Nickel above 1.0 mg/L
Nitrite ion above 0.8 mg/L
Sample pH should be between 3 and 4 to avoid developed color to
fade or turbidity formation.