21
Remove the cuvette and add the content of one packet of
HI 93723-0 reagent. Replace the cap and shake vigorously
for about 10 seconds.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 6 minutes and press Read. When the timer ends the meter will perform the reading. The
instrument displays concentration in
μμ
μμ
μg/L of chromium VI.
Press or to access the second level functions.
Press the Chem Frm key to convert the result in μg/L of Chromate (CrO
4
2
¯) and Dichromate (Cr
2
O
7
2
¯).
Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Vanadium above 1 ppm. However, waiting 10 minutes before reading, the interference is removed.
Iron above 1 ppm.
Mercurous and mercuric ions cause slight inhibition of the reaction.
Chromium VI HR