Hanna-instruments HI 83225 Manual de usuario Pagina 46

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 60
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 45
46
Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
Remove the cuvette and add the content of one
packet of HI 93750B-0 reagent. Replace the cap
and gently mix for one minute by slowly turning the
cuvette upside down.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to
the measurement or, alternatively, wait for 2 minutes and
press Read. When the timer ends the meter will perform the
reading. The instrument displays the results in mg/L (ppm) of
potassium (K).
Press or to access the second level functions.
Press the Chem Frm key to convert the result in mg/L of potassium oxide (K
2
O).
Press or to return to the measurement screen.
INTERFERENCES
Interferences may be caused by:
Ammonium above 10 ppm
Calcium above 10000 ppm as CaCO
3
Chloride above 12000 ppm
Magnesium above 8000 ppm as CaCO
3
Sodium above 8000 ppm
Potassium MR
Vista de pagina 45
1 2 ... 41 42 43 44 45 46 47 48 49 50 51 ... 59 60

Comentarios a estos manuales

Sin comentarios