OUT OF CAL RANGE FUNCTION
The instrument has a mechanism to prevent taking
measurements in a range where the calibration
does not assure the best results. The message
Out of Calibration Range” is displayed on the
message line in the following situations:
• When the first calibration point is over 0.15 NTU and the reading is under 10 NTU.
• When two point calibration was performed and the reading value is over 40 NTU.
When a three point calibration was performed and the reading is over 150% of the the third point value.
When a four point calibration was performed and the reading is over 200% of the the fourth point value.
To take colorimetric measurements follow next
steps:
Turn the instrument on by pressing ON/OFF.
Be sure that the correct range is selected
(displayed on the down-right side of the LCD).
Only the Zero key is available, indicating that
a zero measurement should be done.
ZEROING THE INSTRUMENT
Fill a clean, dry cuvette with 10 mL of sample,
up to the mark, taking care to handle the cuvette
by the top.
Replace the cap.
Wipe the cuvette thoroughly with a lint-free
cloth to remove any fingerprints, dirt or water
spots.
MEASUREMENT PROCEDURE FOR FREE/TOTAL CHLORINE
When taking Free or Total Chlorine measurements, several basic rules should be followed:
Never use cuvettes with scratches or cracks because they can cause inaccurate readings.
Always cap the cuvettes to avoid spillage of the sample into the instrument.
Always close the lid of the instrument during measurement.
If is possible use the same cuvette for zero and sample reading.