Vista de pagina 31
32
Turn the instrument on by toggling ON/OFF switch. When the main screen appears the instrument
is ready.
Make sure that the correct range is selected free or total chlorine. If necessary, press RANGE to
select the correct range.
Press the CAL CHECK key and a screen with
GLP information is displayed including the serial
number and the last calibration time.
Press the Chk key to start the check procedure.
INTERFERENCES
The colorimetric measurements are affected by the following interfering agents:
Bromine (positive error).
Chlorine dioxide (positive error).
Iodine (positive error).
Oxidized Manganese and Chromium (positive error).
Alkalinity above 250 mg/L CaCO
3
or acidity above 150 mg/L CaCO
3
will not reliably develop the
full amount of color or it may rapidly fade. To resolve this, neutralize the sample with diluted
HCl or NaOH.
In case of water with hardness greater than 500 mg/L CaCO
3
, shake the sample for approximatively
2 minutes after adding the powder reagent.
Note:It is recommended to zero the meter before
each measurement.
CAL CHECK PROCEDURE FOR FREE/TOTAL CHLORINE
The HI 83414 free and total chlorine colorimeter has a powerful CAL CHECK function that allows the
user to check the instrument calibration against a NIST traceable standard before making a set of
measurements. With the same standard, the instrument could be re-calibrated, if necessary.
Note:Free and total chlorine must be calibrated separately. Calibration of one range will not
calibrate the other range.
VALIDATION PROCEDURE
Vista de pagina 31